GE LOGIQ E9 Service Manual page 707

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
10-6-8
Isolated patient lead (source) leakage–lead to lead
Select and test each of the ECG lead positions (except ALL) on the LEAD selector, testing each to the
power and ground condition combinations found in:
Table 10-9 on page 10-11
Table 10-10 on page 10-11 as applicable.Record the highest leakage current measured.
5
EV
Chapter 10 Care & Maintenance
LOGIQ E9 Service Manual
10 - 19