7-5-21
Touch Panel
7-5-21-1
Calibration
NOTE:
Always try TP calibration verification before attempting to calibrate.
Verification" on page
To do a Touchscreen Calibration, follow the directions on the Touchscreen. As each of the cross-hairs
appear, touch them with your finger or a pencil eraser.
NOTE:
You MUST hold your finger on the cross-hair until it moves to the next location. If you just tap
the calibration cross-hair, there is a good chance your calibration will be corrupt.
7 - 96
5
EV
7-97.
Section 7-5 - Common Service Desktop
LOGIQ E9 Service Manual
7-5-21-2 "Calibration