Download Print this page

GE LOGIQ E9 Service Manual page 381

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
8-4-6-6
Loading the Application Software (cont'd)
5.) As indicated, press any key to continue.
The process is completely automated and provides you with several status messages, including the
message in
NOTE:
When the LoadSoftware Information Window displays asking you to select OK, do nothing. This
window closes automatically.
6.) Wait while the software loads (approximately 7-8 minutes). Do not interrupt process.
Do not interrupt the software loading at any time.
WARNING
WARNING
After software load is complete, the system reboots.
After the reboot, the System installs the Service Platform.
Do not try to prevent the reboot. If you do, you may end up with hyper threading turned off. If hyper
NOTICE
threading is turned off, the system responds slower than expected.
If hyper threading turns off, you must perform the re-ghost and application software load again.
NOTE:
After the Service Platform installs, and during the second reboot, you may get some error
messages about processes not starting because the workstation is shutting down. If present,
these messages are normal and should be ignored.After the Service Platform reboot, the Start
Application window displays.
Figure 8-11 Start Application window (Note: Window will reflect actual software version instead of "Rx.x.x")
5
EV
Figure 8-10
(LoadSoftware Information Window).
Figure 8-10 LoadSoftware Information Window - Do not select OK
Chapter 8 Replacement procedures
LOGIQ E9 Service Manual
8 - 17

Advertisement

loading