Download Print this page

GE LOGIQ E9 Service Manual page 438

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
8-6-6-1
Upper OP Panel/Touch Panel Assembly removal (cont'd)
Figure 8-36 Upper OP Panel/Touch Panel assembly cable placement
1
Lower OP Panel
2
USB - OUT to Bulkhead Board
3
USB - Video from BEP
4
USB - Upper OP Panel from BEP
NOTE:
The spacers (circled) on the back of the Upper Op Panel are not included with early FRUs
(these will be added to FRUs after product release). So, if you have to swap an Upper Op Panel,
take the spacers off of the old Op panel and swap them into the new one. If they are on so tightly
that you cannot take them off using your fingers, use a pair of pliers and rotate them back and
forth to work them off. These spacers prevent the Op panel from working past its mounting
surface and getting "sucked" too far into the upper frame, causing the Op panel to get twisted
and causing problems.
8 - 74
5
EV
Section 8-6 - Replacing Top Console Parts
5
Power - ON/OFF LOGIC to BEP
6
AN Keyboard USB and Power
7
USB - Trackball
LOGIQ E9 Service Manual