Download Print this page

GE LOGIQ E9 Service Manual page 399

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
8-5-5
Front Cover replacement
8-5-5-1
Manpower
One person, 15 minutes.
8-5-5-2
Tools
For tools needed, refer to:
8-5-5-3
Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the system.
2.) Disconnect the mains power cable from the wall outlet.
3.) Disconnect all Probes and External I/O Cabling.
4.) Remove the Side Covers.
5.) Remove Top Cover.
6.) Remove the Foot Rest Bumper.
Follow these links if you need more information:
4-2-4 "Power shut down" on page 4-7.
8-5-2 "Side Covers replacement" on page 8-27.
8-5-3 "Top Cover replacement" on page 8-31.
8-5-4 "Foot Rest Bumper replacement" on page 8-33.
5
EV
8-2-5 "Tools needed for servicing LOGIQ E9" on page
Chapter 8 Replacement procedures
LOGIQ E9 Service Manual