4-3-9
Contrast checks
For a basic functional check of the system's different modes as well as expected results, refer to the
latest revision of the LOGIQ E9 Basic User Manual, Chapter 5.
4-3-10
Basic Measurements
NOTE:
The following instructions assume that you first scan the patient and then press Freeze.
4-3-10-1
Check Distance and Tissue Depth Measurement
1.) Press Measure once to display an active caliper.
2.) Move the TRACKBALL to position the active caliper at the start point (distance) or the most anterior
point (tissue depth).
3.) Press Set to fix the start point.
4.) The system fixes the first caliper and displays a second active caliper.
5.) Move the TRACKBALL to position the second active caliper at the end point (distance) or the most
posterior point (tissue depth).
6.) Press Set to complete the measurement.
NOTE:
Before you complete a measurement:
To toggle between active calipers, press MEASURE.
To erase the second caliper and the current data measured and start the measurement again, press
CLEAR once.
NOTE:
To rotate through and activate previously fixed calipers, turn Cursor Select.
NOTE:
After you complete the measurement, to erase all data that has been measured to this point,
but not data entered onto worksheets, press Clear.
4 - 38
5
EV
The system displays the distance or tissue depth value in the measurement results window.
Section 4-3 - Functional checks
LOGIQ E9 Service Manual