GE H
EALTHCARE
D
5180263-100, R
IRECTION
Table 8-55
7.
Remove the following parts from the BEP, if present (to reinstall in the replacement BEP):
• DVR Card
• Flex Jumper Cable
• SATA cable (which goes between the DVR and Motherboard)
• Patient I/O filler
• Patient I/O module
• Option Dongle
• Printer Bracket
8 - 134
5
EV
BEP removal - R3.x and earlier
Steps
Section 8-9 - BEP (Back End Processor) parts replacement
LOGIQ E9 Service Manual
Corresponding Graphic