GE LOGIQ E9 Service Manual page 341

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
7-5-21-2
Calibration Verification
The Start Touchscreen Verification brings up a grid. Each time you touch the screen a small red dots
appears where you touch, this will indicate that the calibration is correct and you do not need to
re-calibrate.
If the red dots do not show up at the touched places, you need to run the Touch Panel calibration. To
terminate the verification, move the trackball pointer into the grid and press the set key.
5
EV
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 Service Manual
7 - 97