Download Print this page

GE LOGIQ E9 Service Manual page 285

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
7-5-14
Touchscreen Tab
Test the touchscreen functionality by pressing on the touch screen of the upper operator panel. Verify
that the press is indicated in the black window by a highlighted circle. See: Figure 7-39.
7-5-15
Ending the Program
1.) When you have finished testing the Op Panel, close the GE Test App.
2.) In the Service Platform window, click on the Stop button.
3.) Close the Service Platform window.
4.) Shutdown the system and reboot.
5
EV
Figure 7-39 Touchscreen Tab
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 Service Manual
7 - 41

Advertisement

loading