Download Print this page

GE LOGIQ E9 Service Manual page 337

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
7-5-20-1
Keyboard (cont'd)
Footswitch Tests
Under Service Diagnostics / BEP Interactive/ Keyboard
INPUT > Press left, middle or right footswitch keys.
OUTPUT > Keys will be registered as follows:
-
-
-
5
EV
Left Footswitch = Left Windows + F10
Middle Footswitch = Left Windows + F11
Right Footswitch = Left Windows + F12
Chapter 7 Diagnostics/Troubleshooting