Download Print this page

GE LOGIQ E9 Service Manual page 66

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
2-2-4
Probes environmental requirements
Table 2-6
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
NOTICE
-10 TO +50 degrees C. or +60 degrees C, DEPENDING ON THE TYPE OF PROBE. WHEN
EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE PRODUCT SHOULD BE KEPT IN
ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
Refer to the Table in
2-2-5
Time and manpower requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible, if possible, six weeks
before delivery, to allow enough time to make any changes.
CAUTION
2 - 6
5
EV
Environmental Requirements - Probes
Operation:
Storage:
Temperatures in degrees Celsius (
section 2-2-1-1 on page 2-2
HAVE TWO PEOPLE AVAILABLE TO DELIVER AND UNPACK THE LOGIQ E9.
ATTEMPTS TO MOVE THE LOGIQ E9 CONSIDERABLE DISTANCES OR ON AN INCLINE BY ONE
PERSON COULD RESULT IN INJURY OR DAMAGE OR BOTH.
Section 2-2 - General console requirements
Standard Probes
o
o
10
to 40
C (50 to 104 ºF)
-10° to 60° C (14 to 140 ºF)
-10° to 60° C (14 to 140 ºF)
o
C) conversion to degrees F: (
to determine the needed settlement time.
LOGIQ E9 Service Manual
4D Probes
o
o
18
to 40
C (64.4 to 104 ºF)
o
o
-10
to 50
C (14 to 122 ºF)
o
o
F) = (
C * 9/5) + 32

Advertisement

loading