9-12-3
GTX Boards Compatible Configuration
NOTE:It recommended to replace like for like GTX Boards. If stock of a particular board is not available, use
this table to ensure the correct combination of GTX Boards are installed in the system:
GTX "numbers" use 1, 2, 3 and 4 as nomenclature in VPD, but 0 (zero), 1, 2 and 3 are used as
nomenclature in diagnostics. In this Table, 1, 2, 3 and 4 nomenclature is used to identify board, based
on location in the Card Rack.
Table 9-24
System configuration
Non-CW
CW
MRX - P/N 5393908
5
EV
Card Rack parts
GTX Boards Compatible Configurations
GTX1
(Slot 1)
GRX -
P/N GA200295
MRX -
Not
P/N 5393908-2
populated
MRX -
P/N 5393908-4
GRX -
P/N GA200105
* If replacing P/N 5201044-2 and there is no stock, order P/N 5201044-3 and the Late Request
Plan for FMI 70206 - R2.0.4 Software Upgrade and XYZ USB Cable Update in SN70443A –
Closed FMIs on LOGIQ™ E9, LOGIQ™ 9 and LOGIQ™ 700 Consoles.
** 5201044-3, Rev. D or later are recommended. If P/N 5201044-3 is not available, use
P/N 5201044-5. P/N 5201044-5 can be substituted with P/N 5201044-5.
Chapter 9 Renewal Parts
(cont'd)
GTX1
GTX1
(Slot 2)
(Slot 3)
P/N 5201044-2*
P/N 5201044-2, 5201044-3**, 5201044-5 or
5201044-4
P/N 5201044-3** or 5201044-5
P/N 5201044-3, 5201044-4 or 5201044-5
P/N 5201044-3** or 5201044-5
LOGIQ E9 Service Manual
GTX1
Minimum
(Slot 4)
Software
R 1.0.3
R 1.0.4
R 1.0.5
R 1.0.6
R3.1.1 or later
R3.1.2 or later
R2.x or later
R3.1.2 or later
9 - 51