Launching the Op Panel Test
1.) Log onto the Service Platform, and select the Diagnostics tab. See: Figure 7-29.
2.) Locate the Op Panel Test under I/O Devices > Op Panel Utilities > Op Panel Interface.
3.) Select Start to launch the GE Test App.
An hour glass displays on the Service Platform window. The GE Test App window is hidden behind
the Service Platform window.
4.) When the hour glass displays, minimize the Service Platform window to reveal the GE Test App
window. See:
a.) To minimize, move the cursor to the "-" minus icon at upper right hand corner of the window.
b.) Press the Set key.
5
EV
Figure 7-29 Diagnostics Tab
Figure 7-30 "Select System Type" on page
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 Service Manual
7-32.