Download Print this page

GE LOGIQ E9 Service Manual page 273

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
l
Table 7-1
1.
XYZ Utilities Folder (XYZ Utils)
Access Common Service Desktop as
GE Service. Select Diagnostics tab.
At the end of the list of available Service
Diagnostics, you will find the XYZ Utils
(XYZ Utilities) Folder.
XYZ Utils provides two different tests:
2.
XYZ Version Test
Run the XYZ Version Test.
The system will retrieve the Hardware and
Firmware versions as follows and will show
status PASS.
NOTE: The version displayed in the
illustration may not match the one in the
field. The pass status will indicate that the
system can read the device.
3.
XYZ Fix Park Lock
Run the XYZ Fix Park Lock Test.
The system will release (Undock) the OP
Panel and reset the position of the park
lock. You will hear the motor moving,
releasing the console. The status should
show Passed.
5
EV
Testing the XY Control
Steps
XYZ Version Test
XYZ Fix Park Lock
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 Service Manual
Corresponding Graphic
7 - 29