GE H
EALTHCARE
D
5180263-100, R
IRECTION
8-5-11
Rear Handle replacement
8-5-11-1
Manpower
One person, 15 minutes.
8-5-11-2
Tools
For tools needed, refer to:
8-5-11-3
Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the system
2.) Disconnect the mains power cable from the wall outlet.
3.) Disconnect all Probes and External I/O Cabling.
4.) Remove the Side Covers.
5.) Remove the Rear Cover.
6.) Remove the Top Cover.
Follow these links if you need more information:
4-2-4 "Power shut down" on page 4-7.
8-5-2 "Side Covers replacement" on page 8-27.
8-5-7 "Rear Cover replacement" on page 8-38.
8-5-3 "Top Cover replacement" on page 8-31.
8-5-11-4
Rear Handle removal
1.) Remove the two upper screws (1), one on each side.
2.) Remove the four hexcap screws (2), two on each side.
Figure 8-31 One Phillips and two hexcap screws on each side (left side illustrated)
3.) Lift the Rear Handle away.
5
EV
8-2-5 "Tools needed for servicing LOGIQ E9" on page
Chapter 8 Replacement procedures