IRECTION
Table 3-10
1.
2.
3.
4.
5.
5
EV
Initial Steps - Setup an Image Storage Service
In Utility -> Connectivity -> Service, from the Destination Device drop-down menu, select the
device on which the service resides. (This assumes you have already setup the device in the
Device tab in
3-9-1 "Setup the LOGIQ E9 for DICOM Communications for R2.x.x Software
(some features only apply to R3.x.x)" on page
From the Select Service Type to Add drop-down menu, select DICOM Image Storage and
press Add.
Change the name of the service to one of your choosing.
Enter the AE Title and Port Number of the service. AE Title is case-sensitive.
Save your changes and press Verify. A Smiley Face indicates successful communication with
the service.
NOTE: If you get a successful Ping (Smiley Face) at the device level but not at the service
level, it is possible that the AE Title or Port Number of the service settings are incorrect. Ensure
that these are correct then re-verify. Be sure that the service type (Store, Print, etc) is correct
and supported by the device.
Chapter 3 LOGIQ E9 Setup
LOGIQ E9 Service Manual
3-37).