System Software Load (cont'd)
14.)Remove the base load disk.
NOTE:
If you do not remove the disk, the base load process repeats when the system boots up.
15.)Restart the System using Ctl+Alt+Del.
When the system reboots, it automatically logs on to start checking hardware. This is a normal
process.
NOTE:
If the system does not reboot, press and hold down the Power On/Off button (button turns
amber).
16.)Allow the hardware checking process to run to completion.
NOTE:
Do NOT touch the system during this process. Activating the keyboard, mouse or front
panel could corrupt the installation. While the script is running, several windows or dialog
boxes appear on the screen. At times it may look like the system is unresponsive.
17.)Wait for the "System Settings Change" window asking "Do you want to restart your computer now?"
to display, approximately 1-2 minutes after the desktop is displayed.
Figure 8-7 System Settings Change dialog box - Restart Windows Message
18.)Select "Yes" (see:
NOTE:
In R2.x.x, a DOS window will display, that times out and automatically reboots the system. You
can wait or select "Yes" to restart.
19.)Wait until all drivers are installed automatically and the reboot is complete.
5
EV
Figure
8-7) to restart.
Chapter 8 Replacement procedures
LOGIQ E9 Service Manual
8 - 15