IRECTION
3-9-4-6
How to Setup and Use a DICOM MPPS (Modality Performed Procedure Step) Service
A DICOM Performed Procedure, also known as Modality Performed Procedure Step (MPPS), is a way
to inform your billing system that an ultrasound procedure has been completed and can be billed.
A typical setup has your Worklist Service handle MPPS communications. For example, you can retrieve
today's scheduled exams and report completed exams to the same system.
To set up MPPS on the , follow these steps:
Table 3-21
1.
2.
3.
4.
5.
3 - 62
5
EV
Setup and Use a DICOM MPPS
In Utility -> Connectivity -> Service, from the Destination Device drop-down menu, select
your Worklist device.
From the drop-down menu of services, select Dicom Performed Procedure and press Add.
Provide a name, set the AE Title and Port Number for your Worklist service. AE Title is case-
sensitive.
Maximum Retries, Retry Interval and Timeout have the same use as with DICOM Image
Storage Services. See:
Save your changes and press Verify. A Smiley Face indicates successful communication with
the service.
Section 3-9 - Connectivity Setup and Tips
Utility -> Connectivity -> Service
3-9-4-2 "Properties" on page