IRECTION
10-6-3
Leakage Current Limits
In accordance with these standards, fault conditions like Reverse Polarity of the supply mains and Open
Neutral are no longer required for field evaluation of leakage current. Because the main source of
leakage current is the mains supply, there are different acceptance limits depending on the
configuration of the mains (100-120 or 230-240).
Compare all safety-test results with safety-test results of previously performed safety tests (e.g.
CAUTION
last year etc). In case of unexplainable abrupt changes of safety-test results consult
experienced authorized service personnel or GE for further analysis.
Table 10-9
Chassis/Enclosure Leakage
Type BF and CF Applied Parts
Type CF Applied Parts (sink leakage)
Table 10-10 Leakage current limits for Ultrasound system operation on 230-240 Voltmains
Chassis/Enclosure Leakages
Type BF Applied Parts
Type CF Applied Parts
Type CF Applied Parts (sink leakage)
NOTE:
Leakage current limits for Ultrasound system operation on 230-240 Volt mains
Table 10-11 ISO (on Dale 600) and Mains Applied (on Dale 601) Limits*
BF
CF
*ISO (on Dale 600) and Mains Applied (on Dale 601) refer to the sink leakage test where mains (supply)
voltage is applied to the part to determine the amount of current that will pass (or sink) to ground if a
patient contacted mains voltage.
5
EV
Leakage current limits for Ultrasound system operation on 100-120 Volt mains
(US/Canada/Japan)
Leakage Current Test
Leakage Current Test
Probe Type
5.0 mA
0.05 mA
Chapter 10 Care & Maintenance
System
Lift Ground/
Power
PE Conductor
On and Off
Open
Closed
On (transmit)
Open
On and Off
Closed
System
Lift Ground/
Power
PE Conductor
On
Open and Closed
On (transmit)
Open
On (transmit)
Open
On and Off
Closed
Measurement
LOGIQ E9 Service Manual
Limit mA
0.3
0.01
0.05
0.05
Limit mA
0.5
0.01
0.05
0.05
10 - 11