Download Print this page

GE LOGIQ E9 Service Manual page 475

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
8-6-13
Probe Cable Hooks replacement
8-6-13-1
Manpower
One person, 15 minutes.
8-6-13-2
Tools
For tools needed, refer to:
8-6-13-3
Probe Cable Hook removal
1.) Unscrew the fastening screw.
The screw may be locked with Lock-Tite, so you may need to apply a little extra force.
2.) Remove the Cable Hook.
8-6-13-4
Probe Cable Hook installation
1.) Position the Probe Cable Hook (Figure 8-46).
2.) Fix it in place by using an M5x20 screw with an M6 washer. Lock-Tite has to be used.
Torque: 50 Ncm (0.37 lbf-ft or 4.4 lbf-in) (fingertight).
5
EV
8-2-5 "Tools needed for servicing LOGIQ E9" on page
Figure 8-46 Screw placement, Probe Cable Hook
Chapter 8 Replacement procedures
LOGIQ E9 Service Manual
8-5.
8 - 111

Advertisement