GE H
EALTHCARE
D
5180263-100, R
IRECTION
8-4-6-14
Functional Checks
Perform the functional checks to confirm the system is operational before returning the system to the
customer.
8-4-7
Loading the Application Software Only
Use this procedure to update application software only.
1.) Place the LOGIQ E9 Application Software DVD in the CD/DVD Drive, while the system is running.
2.) Reboot the system. The Start Application window displays.
Figure 8-14 Start Application window (window will reflect actual software version
3.) Select "Install SW." Press OK on the Start Loader message and then press OK on the New System
Software message. A Command window similar to the example shown in
Window" on page 8-22
5
EV
instead of "Rx.x.x")
opens.
Chapter 8 Replacement procedures
LOGIQ E9 Service Manual
Figure 8-15 "Command
8 - 21