3-9-4-8
Structured Reporting
Structured Reporting provides for transmission of exam measurements and calculations in a separate
non-image file. It is not a service in itself, but is a feature of some DICOM Image Storage services, and
is always associated with one. The service must be able to interpret and display the data in the
Structured Report. Refer to the documentation of the Image Storage service documentation for
Structured Reporting.
Table 3-23
To enable Structured Reporting for an Image Storage service, go to the set-up screen for that service
in Utility -> Connectivity -> Service, check the Enable Structured Reporting box and press Save.
From that point on, the sends a Structured Report at the end of an exam with measurements.
3 - 66
5
EV
Structured Reporting
Utility -> Connectivity -> Service - Enable Structured Reporting checked
Section 3-9 - Connectivity Setup and Tips
(continued on next page)
Sending Structure Report
LOGIQ E9 Service Manual