Download Print this page

GE LOGIQ E9 Service Manual page 287

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
7-5-17
Diagnostics for Service Diagnostics
. Also for
7-5-17-1
Digital Receive
See:
7-5-3 "Diagnostics Window Overview" on page 7-24
features.
Symptoms:
Intermittent general problems (communication between DRX and GFI)
Image artifacts - pixilated/jagged edges
1.) Select the Digital Receive checkbox to run all the tests, or select the individual subtests to run only
the selected tests.
2.) Enter the desired Loop Count (numeric value of 4 digits or less).
3.) Click Execute to run the test(s).
5
EV
7-5-21-15 "Diagnostics Window Overview - R3.x.x and later" on page 7-95
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 Service Manual
for a description of the looping and list
7 - 43

Advertisement

loading