EALTHCARE
D
5180263-100, R
IRECTION
7-5-26
Utilities - Scanner Utilities
7-5-26-1
DICOM Verify
This utility provides an easy way to verify DICOM connectivity between the scanner and DICOM devices
on the network.
1.) Enter AE Title, IP Address, and Port values of the DICOM device.
2.) Check the Loop checkbox to repeat the operation, or leave it unchecked to perform the operation
once.
3.) Click the Verify button to see the results.
4.) Uncheck the Loop checkbox to stop the operation.
7 - 116