Download Print this page

GE LOGIQ E9 Service Manual page 209

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
5-2-5
Signal flow overview
Figure 5-3 LOGIQ E9 Signal Flow
GRLY
R
e
l
a
y
s
Weak ultrasound echoes from body structures and blood cells are received by the probes and routed
via the Relay board and the XD bus to the RX board. The RX board amplifies these signals. Then the
signals are routed to the DRX board(s) where the signals are A/D converted. The digital signals are then
further processed on the DRX boards.
5
EV
Front Plane
192
GRX 64
GRX 128
192
192
T/R
TGC
CW
Option
Chapter 5 Components and Functions (Theory)
- GFI Configuration
GTX 64 Ch
GTX 64 Ch
GTX
David
Nathan
Pulser
A/D
0
64 Ch
David
Nathan
1
David
Nathan
2
David
Nathan
Pulser
A/D
3
GDIF
LOGIQ E9 S
DRX 64
DRX 64
DRX
64 Ch
0
GFI
1
2
FDEM
SDEM
ATGC
T_sig
3
GDIF
M
ERVICE
ANUAL
BEP
5 - 7