Download Print this page

GE LOGIQ E9 Service Manual page 485

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
8-7-3-2
Removal Process with Non-loop Type Strap (cont'd)
3.) Using the first and second fingers of your left hand, lever down on the filter removal strap.
NOTE:
This action will prevent your fingers from being struck by the filter when you remove it and it
pops loose.
4.) Remove the filter from the frame.
NOTE:
The filter frame will clear your fingers as you remove the filter.
5
EV
Figure 8-53 Placement of Fingers Around Strap
Figure 8-54 Removing Filter Frame
Chapter 8 Replacement procedures
LOGIQ E9 Service Manual
8 - 121

Advertisement

loading