Loading the Application Software (cont'd)
7.) Remove the applications disk from the DVD drive.
8.) From the Start Applications window, select "Start" (if the CD is still in the tray; otherwise, the system
starts up automatically).
The System powers down and reboots.
The first time the application loads after a SW reload, you may see: the error shown in
"SW Reload Error Message" on page
During the reboot process, the system may prompt for the Basic Option string (valid license) (see:
Figure 8-13 "Enter Option String" on page
If the Basic Option string is not saved in the Option Dongle, or if this is a software upgrade, the SW
License window may open.
NOTE:
If only the title bar is visible in the lower left corner of the monitor, double-click the SW License
title bar to expand the window.
NOTE:
All systems must have at least the Basic Option string for the software to run. Any single Option
String may enable multiple options (i.e. Basic).
NOTE:
If the system does not display an image or a good image when you boot up, Do Not troubleshoot
it as a problem yet. Continue with this install.
9.) Update the Vital Product Data (VPD) information.
Proper Vital Product Data coding is essential to proper system function and optimization. For
information on updating VPD, see:
8 - 18
5
EV
Figure 8-12 SW Reload Error Message
Figure 8-13 Enter Option String
Section 8-4 - Loading / Reloading / Upgrading the Software
8-18. Select OK and reboot the system.
8-18.