Download Print this page

GE LOGIQ E9 Service Manual page 343

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
7-5-22
4D Motor Controller (cont'd)
Name: Static 3D Test
Description: Performs a basic static 3D scan using the RAB2-5 probe connected in probe socket 4.
Run Time: 00:11
If this test failed:
1.) Verify an RAB2-5 probe is connected to probe socket 4 (right-most probe socket).
2.) Verify the correct probe icon appears on the touch panel.
3.) Verify there is no red arrow in the icon indicating that no 4D Motor Controller was detected. Re-run
the 4D Motor Controller Presence Test in this case.
4.) Replace the 4D Motor Controller board.
Name: 4D Test
Description: Performs a basic 4D scan using the RAB2-5 probe connected in probe socket 4.
Run Time: 00:11
If this test failed:
1.) Verify an RAB2-5 probe is connected to probe socket 4 (right-most probe socket).
2.) Verify the correct probe icon appears on the touch panel.
3.) Verify there is no red arrow in the icon indicating that no 4D Motor Controller was detected. Re-run
the 4D Motor Controller Presence Test in this case.
4.) Replace the 4D Motor Controller board.
5
EV
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 Service Manual
7 - 99

Advertisement

loading