Download Print this page

GE LOGIQ E9 Service Manual page 386

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
8-4-7
Loading the Application Software Only (cont'd)
NOTE:
The window reflects the actual software version instead of "R6.0.6EP1_L9_, as shown above.
4.) As indicated, press any key to continue.
NOTE:
When the LoadSoftware Information Window displays asking you to select OK, do nothing. This
window closes automatically.
5.) Wait while the software loads (approximately 7-8 minutes). Do not interrupt this process and ignore
any errors.
Do not interrupt the software load at any time.
WARNING
WARNING
After the software load is complete, the system reboots.
Do not try to prevent the reboot.
NOTICE
8 - 22
5
EV
Figure 8-15 Command Window
1.)
Figure 8-16 LoadSoftware Information Window - Do not select OK
Section 8-4 - Loading / Reloading / Upgrading the Software
LOGIQ E9 Service Manual

Advertisement

loading