Download Print this page

GE LOGIQ E9 Service Manual page 517

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
8-9-5-1
BEP Side I/O Board removal
Table 8-68
1.
Disconnect all I/O cables and all cables at the top of the BEP. See: Figure 8-59 on page 8-137.
2.
Loosen the thumb screws or screws at the top of the BEP cover, or the three screws that attach
the cover to the BEP.
Tilt the top of the BEP cover away from the BEP.
3.
Disconnect:
• the battery cable from PCN1.
• the CB to PB cable from PCN2.
Place the BEP cover aside.
4.
Remove the Option Dongle, on the I/O
Board, inside the BEP.
5
EV
BEP Side I/O Board removal
Steps
Chapter 8 Replacement procedures
LOGIQ E9 Service Manual
Corresponding Graphic
BEP6 CB cables
Option Dongle
8 - 153

Advertisement

loading