BEP replacement - R3.x and earlier (cont'd)
8-9-2-1
BEP removal - R3.x and earlier
Table 8-55
WARNING
WARNING
When you return the used BEP to your local parts organization, make sure you remove ALL
PATIENT DATA from the hard drive, given that the hard drive is still functional.
In some countries, you may be required to delete all software from the disk before returning the
BEP to the parts warehouse. Follow your local policies.
Before you dispose of the hard drive, make sure you remove ALL PATIENT DATA from the hard
drive, given that the hard drive is still functional.
In some countries, you may be required to delete all software from the disk before returning the
hard drive to the parts warehouse. Follow your local policies.
CAUTION
When the BEP door is open, the Extended Power Shutdown circuit board is exposed. When
working inside the BEP, remove the battery cable form J3 on the Extended Power Shutdown.
Otherwise, the 24V battery will be exposed. If you drop a tool or a screw on the Extended Power
Shutdown, you could short the battery. Permanent damage will be done if this happens.
NOTE: Some cables in the BEP have a locking latch. Make sure to slide the locking latch towards you
and pull the connector to the left to release.
1.
Disconnect all I/O cables and all cables at the top of the BEP. See: Figure 8-167 on page 8-383.
2.
Loosen the thumb screw or screws at the top of the BEP cover that attach the cover to the BEP.
Tilt the top of the BEP cover away from the BEP.
3.
Disconnect:
• the battery cable from J3 on the EPS
(Extended Power Shutdown) Board.
• the cables (1) connecting the EPS to the
BEP.
Place the BEP cover aside.
8 - 132
5
EV
BEP removal - R3.x and earlier
Steps
Section 8-9 - BEP (Back End Processor) parts replacement
LOGIQ E9 Service Manual
Corresponding Graphic
inside of BEP Cover, EPS and cables