Table 8-16
3.
WHEN REMOVING THE LCD FROM THE
ARM, KEEP THE LCD ARM ASSEMBLY
IN THE LOCKED POSITION. THE
SPRINGS TO SUPPORT THE LCD CAN
CAUSE THE ARM TO SPRING OPEN
CAUSING SEVERE PERSONAL INJURY
AND PROPERTY DAMAGE.
4.
Keep the LCD supported and remove the
two M5x20 screws and spring "lock"
washers which mount the LCD to the Arm
Assembly.
5
EV
LCD Monitor V2 removal (used in R4.x and later production)