Download Print this page

GE LOGIQ E9 Service Manual page 478

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
8-7-2
Rear Filter and "handle type" Bottom Filter replacement / cleaning
8-7-2-1
Manpower
One person, 15 minutes.
8-7-2-2
Tools
No tools are needed to replace the filter.
8-7-2-3
Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the system.
2.) Disconnect the mains power cable from the wall outlet.
3.) Remove the Filter Cover.
Follow these links if you need more information:
4-2-4 "Power shut down" on page 4-7.
8-5-6 "Filter Cover replacement" on page 8-37.
Clean the air filters to ensure that a clogged filter does not cause the system to overheat and reduce
system performance and reliability. It is recommended the filters be cleaned quarterly (once every three
months) or more often in locations where high levels of dust are present.
The LOGIQ E9 has two air filters which need to be cleaned. The top air filter is located on the back of
the LOGIQ E9 below the power cord and the bottom air filter is located underneath the LOGIQ E9.
Table 8-52
Power down the system before
1.
removing the filters to prevent any
loose or knocked-off debris from
entering the Fan Tray.
Walk the LOGIQ E9 forward until the
caster is in position to access the filter
handle. The right side, rear caster
must be in-line and away from the
LOGIQ E9.
8 - 114
5
EV
Removing and Cleaning Filters
Steps
Section 8-7 - Main Console parts replacement
LOGIQ E9 Service Manual
Corresponding Graphic

Advertisement

loading