Download Print this page

GE LOGIQ E9 Service Manual page 556

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
8-10-4-4-3
4D MC Replacement for Lambda PS
.
Table 8-96
1.
Locate the 4D MC Cover on the back of
the Main PS.
Remove the three screws from the
4D MC Cover. DO NOT discard. Gently
slide the 4D MC Cover and the 4D MC out
of the PS.
Remove the two screws (circled) that
2.
secure the 4D MC to the 4D MC Cover.
Discard. Note how the 4D MC USB Cable
is routed for installation.
Disconnect the USB Cable from J28 on
3.
the BEP.
8 - 192
5
EV
4D MC Replacement for Lambda PS
Steps
Section 8-10 - Main Power Supply parts replacement
LOGIQ E9 Service Manual
Corresponding Graphic

Advertisement

loading