EALTHCARE
D
5180263-100, R
IRECTION
Section 6-5
Direction Lock and Brake adjustments
There are no adjustments for the Direction Lock or the brakes.
Section 6-6
Adjust time-out for DICOM servers
When portable (off-line), use minimum time-out and no retries or it will affect shutdown speed.
The absolute value of the magnetic distortion reading ("1-2 Distortion") should be consistently less than
1 mm, especially since the transducer is held steady in one location. It is also important that the reading
is not cycling quickly between values such as continuously looping from -0.2 to 0.6. The environmental
quality for both sensor 1 and sensor 2 should be in the green range, preferably at 6 or 7 bars.