GE H
EALTHCARE
D
5180263-100, R
IRECTION
8-9-2-2
BEP installation - R3.x and earlier
Table 8-56
1.
Reinstall the following parts from the BEP that was replaced (all that were present):
• DVR Card
• Flex Jumper Cable
• SATA cable (which goes between the DVR and Motherboard)
• Patient I/O filler
• Patient I/O module
• Option Dongle
• Printer Bracket
2.
Slide the BEP into the left side of the chassis frame.
3.
Install the two HEX key screws at the inside
base of the BEP (Torque: 10 Nm
{7.4 lbf-ft}).
5
EV
BEP installation - R3.x and earlier
Steps
Chapter 8 Replacement procedures
LOGIQ E9 Service Manual