GE LOGIQ E9 Service Manual page 54

Hide thumbs Also See for LOGIQ E9:

Advertisement

GE H
EALTHCARE
D
5180263-100, R
IRECTION
1-4-4
Electrical safety
1-4-4-1
Safe practices
Follow these guidelines to minimize shock hazards whenever you are using the scanner:
The equipment chassis must be connected to an electrical ground.
The unit is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety ground. A separate power outlet with a 20 amp circuit breaker for 120
VAC for 120V area, 7.5 amp circuit breaker for 220-240 VAC for 220/240V area or 15 amp circuit
breaker for 100 VAC for Japan.
The power outlet used for this equipment should not be shared with other types of equipment.
Both the system power cable and the power connector must meet international electrical standards.
Connecting a LOGIQ E9 scanner to the wrong voltage level will most likely destroy it.
WARNING
WARNING
1-4-4-2
Probes
Follow these guidelines before connecting a probe to the scanner:
Inspect the probe prior to each use for damage or degradation to the:
-
-
-
-
-
-
Do not use a damaged or defective probe.
Never immerse the probe connector or adapter into any liquid.
The system has more than one type of probe port. Use the appropriate probe port designed for the
probe you are connecting.
Section 1-5
Label locations
Refer to the current revision of the LOGIQ E9 Basic User Manual, 5180374-100, Chapter 2.
1 - 20
5
EV
housing
cable strain relief
lens
seal
connector pins
locking mechanism
Section 1-5 - Label locations
LOGIQ E9 Service Manual