11. Replace the printer bracket at the top of the BEP.
Be sure the lip, on the underside of the bracket, hooks on the edge of the card rack, and the three
tabs insert into the slots on the top of the BEP frame. The lip "clamps" the card rack and BEP
together. This is a tight fit.
12. Lower the lower Column Cover. Make sure the Column Cover is behind the printer bracket.
13. Position the Front Cover to engage the Column Cover stop tabs.
14. Latch the two latches that clamp the printer bracket to the top of the BEP.
15. Slide the Black and White printer into the
printer bracket and connect the cables to
the back of the printer.
16. Tighten the printer bracket wing nut to secure the printer.
17. Connect all cables at the top of the BEP and all I/O cables .
18. Replace the covers.
19. Reinstall Baseload and Application software.
20. Enter the system's TCPIP settings and Option strings.
21. Restore the Patient Archive images, Report Archive, User-Defined Configuration (Customer
Presets), and Service settings.
22. Perform Functional Checks. See:
and