10-6-9-1
Isolated lead (sink) leakage test record
Table 10-14 on page 10-22shows a typical format for recording the isolated patient lead sink leakage
current.
Measurements should be recorded for the full lead combination under each set of test conditions
specified in:
Table 10-9 on page 10-11
Table 10-10 on page 10-11
Record all data on the Inspection Certificate.
Table 10-14 Typical data format for recording isolated lead (sink) leakage
NOTE:
Values in italics font are given as examples only.
1.) Connect Safety analyzer to wall AC power outlet
2.) Plug the equipment under test power cable into the receptacle on the panel of the meter.
3.) Connect the ECG cable to the scanner and the Patient leads to the analyzer
4.) Select the Lead isolation or main applied function on the meter
NOTE:
Consult the manufacturer's user manual of the Safety Analyzer
Test opening and closing the ground with the scanner on and off as indicated in Table 10-9 or
Table 10-10
NOTE:
Consult the manufacturer's user manual of the Safety Analyzer
10 - 22
. 2
EV
as applicable.
Unit under test___________________________ Date of test:_____________________
Test Conditions
System Power
On
Off
as applicable
Section 10-6 - Electrical Safety Tests
Patient Lead
Grounding/PE
Closed
Closed
LOGIQ E9 S
M
ERVICE
ANUAL