GE LOGIQ E9 Service Manual page 367

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
8-4-6-6
Loading the Application Software (cont'd)
7.) Remove the applications disk from the DVD drive.
8.) From the Start Applications window, select "Start" (if the CD is still in the tray; otherwise, the system
starts up automatically).
The System powers down and reboots.
The first time the application loads after a SW reload, you may see: the error shown in
Select OK and reboot the system.
During the reboot process, the system may prompt for the Basic Option string (valid license) (see:
Figure 8-13 "Enter Option String" on page
If the Basic Option string is not saved in the Option Dongle, or if this is a software upgrade, the SW
License window may open.
NOTE:
If only the title bar is visible in the lower left corner of the monitor, double-click the SW License
title bar to expand the window.
NOTE:
All systems must have at least the Basic Option string for the software to run. Any single Option
String may enable multiple options (i.e. Basic).
Demo Option strings turn on all options for a limited period of time. Do Not leave a demo Option
string in a unit when it is sold to a customer.
8-4-6-7
Verifying the software versions
1.) From the Touch Panel, select Utility -> System -> ABOUT.
2.) Verify software and hardware versions as recorded in
versions" on page
. 2
EV
Figure 8-12 SW Reload Error Message
Figure 8-13 Enter Option String
4-30.
Chapter 8 Replacement procedures
8-21.
Table 4-14 "Record Software and Hardware
LOGIQ E9 S
M
ERVICE
ANUAL
Figure
8-12.
8 - 21

Advertisement

Table of Contents
loading

Table of Contents