GE LOGIQ E9 Service Manual page 209

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
5-2-7
Signal flow overview (cont'd)
Figure 5-6 LOGIQ E9 Signal Flow
GRLY
R
256
e
l
a
y
s
The MRX Board combines the functionality of the Receiver Boards (DRX and GRX) and the GFI.Weak
ultrasound echoes from body structures and blood cells are received by the probes and routed via the
Relay board and the XD bus to the MRX board. The MRX board amplifies the ultrasound signal and
connects it with an A/D converter to the digital domain. The digital signals are then further processed
on the MRX boards.
. 2
EV
Frontplane
MRX
MRX 192 Ch
TGC
TGC
TGC
TGC
T/ R
TGC
Chapter 5 Components and Functions (Theory)
- MRX
256
CW
Option
A/D
Converters
I / Q
Filter
X
I
Analog
Beamformer
X
Q
X
A/DA
Nathan
A/DA
32 Ch
Nathan
A/DA
32 Ch
A/DA
Nathan
32 Ch
Nathan
32 Ch
Nathan
32 Ch
Nathan
32 Ch
A/D
LOGIQ E9 S
ERVICE
GTX 64 Ch
GTX 64 Ch
GTX
David
Pulser
64 Ch
0
David
1
David
2
David
Pulser
3
GDIF
Front End
Interface
DSP
SED
TGC
Test
GFE
Gen
Gen
M
ANUAL
5 - 11

Advertisement

Table of Contents
loading

Table of Contents