5535208-100, R
IRECTION
Table 8-114 Removing the Fan Tray Filter
2.
Flip the Fan Tray over on a clean, safe
surface.
Pull out the Filter latch and slide the Filter
out.
Clean the Bottom Filter after it is removed
by removing excess lint or dust from the
soiled side. If necessary, use a clean, soft
brush; or vacuum.
3.
Reinstall the Filter into the Fan Tray and reinstall the Fan Tray.
completely into the Card Cage.
8 - 188
. 2
EV
Steps
Section 8-7 - Main Console parts replacement
LOGIQ E9 S
ERVICE
Corresponding Graphic
Make sure the Fan Tray seats
M
ANUAL

Advertisement

Table of Contents