Section 6-5

Direction Lock and Brake adjustments

There are no adjustments for the Direction Lock or the brakes.
Section 6-6

Adjust time-out for DICOM servers

When portable (off-line), use minimum time-out and no retries or it will affect shutdown speed.
The absolute value of the magnetic distortion reading ("1-2 Distortion") should be consistently less than
1 mm, especially since the transducer is held steady in one location. It is also important that the reading
is not cycling quickly between values such as continuously looping from -0.2 to 0.6. The environmental
quality for both sensor 1 and sensor 2 should be in the green range, preferably at 6 or 7 bars.
6 - 16
. 2
EV
Section 6-5 - Direction Lock and Brake adjustments
LOGIQ E9 S
M
ERVICE
ANUAL