3D/4D And Volume Navigation Functional Check; Swave (Shear Wave "Shear Elasto") Functional Check - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
Section 4-6

3D/4D and Volume Navigation Functional Check

For complete information, refer to the appropriate version of the LOGIQ E9 Basic User Manual,
Chapters 5 and 6.
6-10-5 "Setup" on page 6-306-10-6 "Measure Initial Accuracy" on page 6-31
Section 4-7

SWAVE (Shear Wave "Shear Elasto") Functional Check

If the Capacitor Pack is not present or the Power Supply to Capacitor Pack Cable is not connected, or
connected properly, the option to select the "Shear Elasto" soft button after pressing the Elasto button,
the Touch Panel key "Shear Elasto/strain Elasto" button WILL NOT be present.
Select a probe (9L D and C1-6D).
Select ELASTO button.
In the Touch Panel, select Shear Wave button if not already selected.
Verify that the Shear Wave button is available.