Chapter 2
Site preparations
Section 2-1

Overview

2-1-1
Purpose of this chapter
This chapter provides the information required to plan and prepare for the setup of a LOGIQ E9.
Included are descriptions of the facility and electrical needs to be met by the purchaser of the LOGIQ E9.
Section 2-2
General console requirements
2-2-1
Console environmental requirements
2-2-1-1
If the LOGIQ E9 is very cold or hot
When unpacking the device, allow the temperature of the device to stabilize before powering up. The
following table describes guidelines for reaching operational temperatures from storage or transport
temperatures. See:
Table 2-1
o
60
55
50
C
o
140
131
122
113
F
Hrs
8
6
4
2-2-1-2
Environmental specifications
Table 2-2
Temperature
Humidity
Pressure
Heat Dissipation
. 2
EV
Table
2-1'
LOGIQ E9 Acclimate Time
45
40
35
30
25
104
95
86
77
2
0
0
0
0
Environmental Specifications for LOGIQ E9 Scanners
Operational
10º - 35º C
(50º - 96º F)
30% - 80%
non-condensing
70 - 106 kPa
4712 Btu/hour
Chapter 2 Site preparations
20
15
10
5
0
68
59
50
41
32
0
0
0
2
4
Storage
-10º - 50º C
(-14º - 122º F)
30% - 80%
non-condensing
70 - 106 kPa
"
LOGIQ E9 S
-5
-10
-15
-20
-25
23
14
5
-4
-13
6
8
10
12
14
Transport (< 16 hrs.)
-10º - 50º C
(-14º - 122º F)
30% - 80%
non-condensing
70 - 106 kPa
M
ERVICE
ANUAL
-30
-35
-40
-22
-31
-40
16
18
20
"
2 - 1