GE LOGIQ E9 Service Manual page 587

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
8-9-9-1
Battery Pack replacement
Table 8-153 Battery Pack removal
1.
Loosen the thumb screws or screws at the top of the BEP cover, or the three screws that attach
the cover to the BEP.
Tilt the top of the BEP cover away from the BEP.
2.
Disconnect the battery from PCN1 on the
the CB (ChargeBoard).
Position the cable connector into the
opening of the battery cavity, so the cable
and battery can be easily removed.
Close BEP door and secure closed.
3.
Remove the four screws to the Battery
Cover on the BEP cover.
Remove the Battery Pack Cover.
4.
Remove the stabilizing foam strip beside the battery pack.
5.
Remove the BEP Battery Pack.
. 2
EV
Steps
Chapter 8 Replacement procedures
LOGIQ E9 S
Corresponding Graphic
BEP6 CB cables
BEP Cover, BEP Battery Pack Cover screws
M