Table of Contents

Advertisement

Quick Links

P R E L I M I N A R Y
GE
Voluson i / Voluson e
Service Manual
• Voluson i / Voluson e systems with Software version SW 6.x.x (BT06)
• Voluson i / Voluson e systems with Software version SW 7.x.x (BT07)
• Voluson i / Voluson e systems with Software version SW 8.x.x (BT09/11)
• Voluson i / Voluson e systems with Software version SW CC200 (BT14)
(NOVEMBER 8, 2013)
VE8_COVER.FM
Part Number: KTI106052
Revision: 10
0123

Advertisement

Table of Contents
loading

Summary of Contents for GE Voluson i BT06

  • Page 1 P R E L I M I N A R Y (NOVEMBER 8, 2013) VE8_COVER.FM Voluson i / Voluson e Service Manual • Voluson i / Voluson e systems with Software version SW 6.x.x (BT06) • Voluson i / Voluson e systems with Software version SW 7.x.x (BT07) •...
  • Page 3 • VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES WARNUNG KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN (DE) WURDE. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
  • Page 4 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
  • Page 5 ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝΔΥΝΟΥΣ. EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ EL. • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A FIGYELMEZTETÉS KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
  • Page 6 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU. • REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG VIÐVÖRUN SKILIÐ...
  • Page 7 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ. • JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU. • NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS BRĪDINĀJUMS IZLASĪŠANAS UN SAPRAŠANAS.
  • Page 8 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ. • DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O TRADUCERE.
  • Page 9 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL TA SERVISNI PRIROČNIK JE NA VOLJO SAMO V ANGLEŠČINI. • ČE PONUDNIK SERVISNIH STORITEV ZA STRANKO POTREBUJE NAVODILA V DRUGEM JEZIKU, JE ZA PREVOD ODGOVORNA STRANKA SAMA. • NE POSKUŠAJTE SERVISIRATI OPREME, NE DA BI PREJ PREBRALI IN OPOZORILO RAZUMELI SERVISNI PRIROČNIK.
  • Page 10 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR. • EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ BİR DİLDE OLMASINI İSTERSE, KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN SORUMLULUĞUNDADIR. DİKKAT • SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE (TR) ETMEYİNİZ.
  • Page 11 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL (ZH-CN) (KO)
  • Page 12 GE personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 13 The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE. GE may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
  • Page 14 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Revision History Revision Date Reason for change April 14, 2006 Initial Release August 18, 2006 Implementation of new parts and peripherals, general update and modification November 17, 2006 Introduction of Voluson® e, software version 6.1.x April 04, 2007 Implementation of new parts, release of BT´07 software version 7.0.x October 19, 2007...
  • Page 15: Table Of Contents

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
  • Page 16 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Contact Information ......... . . 1 - 18 System Manufacturer .
  • Page 17 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL CHAPTER 2 Site-Preparation Overview............2 - 1 Purpose of Chapter 2 .
  • Page 18 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL CHAPTER 3 Setup Instructions Overview ............3 - 1 The Purpose of Chapter 3 .
  • Page 19 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Connection and Usage of the optional Voluson Station ... . . 3 - 35 Attaching the Voluson i / Voluson e to the Voluson Station ..3 - 35 Installation Gel Condom Holder / Probe Holder for Endocavity Probe 3 - 37 Installation of the Additional Shelf Peripherals .
  • Page 20 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Voluson i / Voluson e mounted on Dock Cart ....3 - 51 Voluson i / Voluson e mounted on Voluson Station .
  • Page 21 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Adjustment of Printer Settings ........3 - 91 UP-D897 - Printer Settings .
  • Page 22 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Dicom®/SonoView Configuration ....... . . 3 - 117 Map Network Drive .
  • Page 23 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL CHAPTER 4 Functional Checks Overview ............4 - 1 Purpose of Chapter 4 .
  • Page 24 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Deactivating the probe ........4 - 25 Disconnecting the probe .
  • Page 25 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview............5 - 1 Purpose of Chapter 5 .
  • Page 26 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL SonoNT ..........5 - 17 Sono Render Start .
  • Page 27 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Keyboard and Control Panel Components ......5 - 39 Power ON/OFF Button .
  • Page 28 GE-Service ........
  • Page 29 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL CHAPTER 6 Service Adjustments Overview............6 - 1 Purpose of Chapter 6 .
  • Page 30 No Connection to the Network at All ......7 - 26 GE remote service connection ......7 - 26...
  • Page 31 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Error Messages ........... 7 - 27 Table of Contents xxix...
  • Page 32 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview ............8 - 1 Purpose of Chapter 8 .
  • Page 33 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Preparations ..........8 - 26 TGC Slider - Removal Procedure .
  • Page 34 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Rigid Flex Assy - Installation Procedure ......8 - 34 Replacement of the LCD Backlight Inverter Board .
  • Page 35 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Preparations ......... . . 8 - 43 Wheels - Removal Procedure .
  • Page 36 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Fuses - Installation Procedure ......8 - 51 Replacement of the Wheels .
  • Page 37 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Manpower ..........8 - 61 Tools .
  • Page 38 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Probe Mux Box - Installation Procedure ..... . 8 - 69 Replacement of the Patient Monitor .
  • Page 39 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Preparations ......... . . 8 - 77 Top Column Front Cover - Removal Procedure .
  • Page 40 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Manpower ..........8 - 86 Tools .
  • Page 41 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL CHAPTER 9 Renewal Parts Overview............9 - 1 Purpose of Chapter 9 .
  • Page 42 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL CHAPTER 10 Care & Maintenance Overview ............10 - 1 Periodic Maintenance Inspections .
  • Page 43 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Meter Procedure ........10 - 17 Chassis Leakage Current Test .
  • Page 44 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL xlii Table of Contents...
  • Page 45: Overview

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the Voluson i / Voluson e. The service provider must read and understand all the information presented in this manual before installing or servicing a unit.
  • Page 46: Typical Users Of The Basic Service Manual

    IRECTION EVISION ERVICE ANUAL 1-1-3 Typical Users of the Basic Service Manual • GE Service Personnel (installation, maintenance, etc.). • Hospital’s Service Personnel • Contractors (Some parts of Chapter 2 - Site Preparation) 1-1-4 Models Covered by this Manual Table 1-2...
  • Page 47: Purpose Of Operator Manual(S)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 1-1-6 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the Voluson i / Voluson e and also kept near the unit for quick reference. Section 1-2 Important Conventions 1-2-1...
  • Page 48: Standard Hazard Icons

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
  • Page 49: Product Labels And Icons

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 1-2-3 Product Labels and Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Description of symbols and labels used in combination with this Voluson ultrasound system. Product labels and its location depend on BT-version and/or date of manufacture.
  • Page 50 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION “CAUTION This unit weighs... This precaution is intended to prevent injury that may result if Special care must be used to avoid..." one person attempt to move heavy parts (e.g., the optional Used in the Service and User Manual Modo Cart, Dock Cart or Voluson Station) considerable...
  • Page 51 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION "Protective Earth" Indicates the protective earth (grounding) various terminal. "Equipotentiality" Indicates the terminal to be used for peripherals connecting equipotential conductors when interconnecting isolation transformer on Modo Cart (grounding) with other equipment.
  • Page 52: Safety Considerations

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
  • Page 53: Electrical Safety

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 1-3-3 Mechanical Safety (cont’d) Always: • Eject any DVD/CD from the drive. • Ensure that the Voluson i / Voluson e system is well prepared and packed in its original packaging before transporting.
  • Page 54: Auxiliary Devices Safety

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 1-3-5 Auxiliary Devices Safety Power Supplies for additional equipment MUST comply with IEC 60601-1. WARNING WARNING !! WARNING: WARNING WARNING DO NOT attempt to use different peripherals and accessories (brand and model; connected via the PCMCIA or USB ports) other than approved and provided by !! WARNING: The ultrasound system is an extremely sensitive and complex medical system.
  • Page 55: Label Location

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 1-3-5 Auxiliary Devices Safety (cont’d) Please observe that some printers may not be medical devices! If the Bluetooth Printer and/or CAUTION Line Printers are no medical devices, they have to be located outside of the patient environment !! CAUTION: (according to IEC 60601-1 / UL 60601-1).
  • Page 56: Main Label On Voluson I / Voluson E System

    6 Safety type: Type BF 9 GHOST R label 1-3-6-2 Main Label on optional Modo Cart The Main Label is located on the back of the Modo Cart. GE Medical Systems 7S126_2 Kretztechnik GmbH & Co OHG Tiefenbach 15 4871 Zipf Austria...
  • Page 57: Main Label On Optional Dock Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 1-3-6-3 Main Label on optional Dock Cart The Main Label is located on the back of the Dock Cart. Figure 1-4 Main Label (located on back of the Dock Cart) 1 Model Type 4 Manufacturer 2 Input Specification...
  • Page 58: Dangerous Procedure Warnings

    NOTE: Regulations vary for different countries. Dispose of a used battery in accordance with local regulations. Use only batteries approved by GE as suitable for use with the Voluson i / Voluson e portable CAUTION ultrasound system.
  • Page 59 If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal during use, recharging or storage, immediately remove it and stop using it. • If you have any questions about the battery, consult your local GE representative. • Storage of the battery pack: Short term (less than 1 month): between 0°...
  • Page 60: Lockout/Tagout Requirements (For Usa Only)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 1-3-9 Lockout/Tagout Requirements (For USA Only) Follow OSHA Lockout/Tagout requirements to protect service personnel from injuries caused by unexpected energizing or start-up of equipment during service, repair, or maintenance. Energy Control and Power Lockout for Voluson i / Voluson e.
  • Page 61: Electromagnetic Compatibility (Emc)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 1-4 Electromagnetic Compatibility (EMC) 1-4-1 What is EMC? Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface.
  • Page 62: Customer Assistance

    Korea +(82) 2620 13585 Singapore +(95) 6277-3444 1-5-2 System Manufacturer Table 1-7 System Manufacturer Manufacturer Telephone GE Healthcare Austria GmbH & Co OG Tiefenbach 15 +43 (0) 7682-3800-0 +43 (0) 7682-3800-47 A-4871 Zipf Austria 1-18 Section 1-5 - Customer Assistance...
  • Page 63: Site-Preparation

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Chapter 2 Site-Preparation Section 2-1 Overview 2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a Voluson i / Voluson e ultrasound unit. Included are descriptions of the facility and electrical needs to be met by the purchaser.
  • Page 64: Electrical Requirements

    The Voluson i / Voluson e portable ultrasound system is supplied with a main power plug, as standard. In the event that the unit arrives without a power plug, or with the wrong plug, contact your GE dealer. When necessary, the installation engineer will supply the appropriate power plug to meet the applicable local regulations.
  • Page 65: Emi Limitations

    RF gaskets touch metal. appropriate location. The interconnect cables are grounded and require ferrite beads and other shielding. Use GE-specified harnesses Cable length, material, and routing are all important; do not make any changes that do not and peripherals. meet all specifications.
  • Page 66: Probe Environmental Requirements

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 2-2-4 Probe Environmental Requirements Operation: Ambient temperature 18° to 30° C Storage: -10° to 50° C NOTE: Temperature in degrees C. Conversion to degrees F = °C (9/5) + 32). SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF NOTICE -10 TO + 50 degrees C.
  • Page 67: Facility Needs

    (and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
  • Page 68: Mandatory Site Requirements

    NOTE: All relevant preliminary network port installations at the prepared site must be performed by authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to perform servicing of the equipment. 2-3-3 Site Recommendations The following are (optional) site recommendations.
  • Page 69: Recommended Ultrasound Room Layout

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 2-3-3-1 Recommended Ultrasound Room Layout Figure 2-1 shows a floor plan illustrating the recommended layout of the Ultrasound Room and depicting the minimal room layout requirements. Dedicated Power Outlets VOLUSON i Hospital Network Cabinet for...
  • Page 70: Networking Setup Requirements

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 2-3-4 Networking Setup Requirements 2-3-4-1 Stand-alone Unit (without Network Connection) None 2-3-4-2 Unit Connected to Hospital’s Network Supported networks: • Ethernet • Wireless LAN 2-3-4-3 Purpose of the DICOM® Network Function DICOM®...
  • Page 71: Setup Instructions

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Chapter 3 Setup Instructions Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information needed to setup the Voluson i / Voluson e ultrasound unit. Included are procedures to receive, unpack and configure the equipment. NOTE: Depending on the customer’s specific requirements, the Voluson i / Voluson e may have been supplied with or without the Voluson Modo Cart, Dock Cart or Voluson Station (optional).
  • Page 72: Set Up Reminders

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 3-2 Set Up Reminders 3-2-1 Average Installation Time The Voluson i / Voluson e installation and functional checkout will take approximately one hour; Voluson i / Voluson e consoles with optional equipment may take slightly longer. Once the site has been prepared, the average installation time required is shown in Table 3-2 below.
  • Page 73: Safety Reminders

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-2-3 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE DANGER AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT !! DANGER: TOUCH THE UNIT! Two people are required to unpack the optional Cart because of its weight.
  • Page 74: Receiving And Unpacking The Equipment

    Unpack the devices such a way that packaging can be reused. For warranty purposes, storage of the above is required for one year from date of purchase. If the shipping crate is damaged, please inform the GE sales representative immediately. NOTICE...
  • Page 75: Unpacking Procedure

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont’d) 3-3-1 Unpacking Procedure 1.) Loosen the screws and then remove the top cover from the crate. top cover of crate Wooden Base Figure 3-3 Shipping crate - Voluson i / Voluson e with Peripherals When using sharp tools to open packing materials, take care to avoid cutting or damaging any CAUTION...
  • Page 76: Unpacking The Voluson I / Voluson E Modo Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-3-2 Unpacking the Voluson i / Voluson e Modo Cart 1.) Cut the straps securing the shipping carton to the wooden base. When using sharp tools to open packing materials, take care to avoid cutting or damaging any CAUTION of the contents.
  • Page 77: Unpacking The Voluson I / Voluson E Dock Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-3-3 Unpacking the Voluson i / Voluson e Dock Cart 1.) Cut the straps securing the shipping carton to the wooden base. When using sharp tools to open packing materials, take care to avoid cutting or damaging any CAUTION of the contents.
  • Page 78: Unpacking The Voluson I / Voluson E Voluson Station

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-3-4 Unpacking the Voluson i / Voluson e Voluson Station 1.) Cut the straps securing the shipping carton to the wooden base. When using sharp tools to open packing materials, take care to avoid cutting or damaging any CAUTION of the contents.
  • Page 79: Preparing For Set Up

    Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any deterioration. Power Cord Verify continuity. Replace the power cord, as required. NOTE: Report any items that are missing, back-ordered, or damaged, to your GE sales representative. The contact address is shown in Contact Information on page 1-18.
  • Page 80 Verify that the voltage of the optional Modo Cart is correct. Voltage see: Section 3-4-3 "Modo Cart Voltage Configuration" on page 3-20 NOTE: Report any items that are missing, back-ordered, or damaged, to your GE sales representative. The contact address is shown in Contact Information on page 1-18. •...
  • Page 81 Section 3-4-4 "Dock Cart / Dock Cart light / Voluson Station Voltage Configuration" on page 3-21 NOTE: Report any items that are missing, back-ordered, or damaged, to your GE sales representative. The contact address is shown in Contact Information on page 1-18.
  • Page 82: Front And Side View Of The Voluson I / Voluson E System

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-4-1-1 Front and Side View of the Voluson i / Voluson e System Figure 3-8 below shows the Voluson i / Voluson e portable ultrasound system components that are visible from the front and side of the ultrasound unit.
  • Page 83: Rear View Of The Voluson I / Voluson E System

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-4-1-2 Rear View of the Voluson i / Voluson e System Figure 3-9 shows the view of the Voluson i / Voluson e ultrasound unit rear panel and external peripheral/accessory connectors: Figure 3-9 View of the Voluson i / Voluson e Rear Panel Two interchangeable USB ports (Printers, DVD+R/RW Drive and other...
  • Page 84: Front View Of The Modo Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-4-1-3 Front View of the Modo Cart Figure 3-10 Modo Cart - Front View Changeable probe/gel holders Moving handle Footrest Upper peripheral storage shelf (e.g., for DVD+R/RW Drive) Middle peripheral storage shelf (e.g., for Color printer, DVD recorder, ...) Lower peripheral storage shelf (e.g., for B/W printer) Front Caster wheels Brakes...
  • Page 85: Rear View Of The Modo Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-4-1-4 Rear View of the Modo Cart Figure 3-11 Modo Cart - Rear View Cable hooks Power outlets for Peripherals Isolation Transformer Fuses for changing between 115V and 230V Mains Power Switch Power Cord + pull out protection mounted Screws for changing the height of the mounting platform...
  • Page 86: Front View Of The Dock Cart / Dock Cart Light

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-4-1-5 Front View of the Dock Cart / Dock Cart light Figure 3-12 Dock Cart/Dock Cart light - Front View Tray Recorder USB Hub (not available on Dock Cart light) Changeable probe/gel holders Moving handle Upper peripheral storage shelf (e.g., for Color printer)
  • Page 87: Rear View Of The Dock Cart / Dock Cart Light

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-4-1-6 Rear View of the Dock Cart / Dock Cart light Figure 3-13 Dock Cart/Dock Cart light - Rear View Lan Isolation Box (not available on Dock Cart light) Moving Handles Release Lever (not available on Dock Cart light) Cable Guide...
  • Page 88: Front View Of The Voluson Station

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-4-1-7 Front View of the Voluson Station Figure 3-14 Voluson Station - Front View Patient Monitor Changeable probe/gel holders optional Probe Mux Box Cover BW Printer and Console Bottom Shelf Footrest Brakes Front Caster Wheels...
  • Page 89: Rear View Of The Voluson Station

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-4-1-8 Rear View of the Voluson Station Figure 3-15 Voluson Station - Rear View Release Lever Changeable probe/gel holders Additional Shelf Peripherals (e.g., for Color printer) Storage Tray Back Mains Power Switch Power Supply Fuses for changing between 115V and 230V Voltage Output Selector switch (115V and 230V)
  • Page 90: System Voltage Settings

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-4-2 System Voltage Settings NOTE: Not necessary at the Voluson i / Voluson e portable ultrasound system: For details see: Table 2-3 "Electrical Specifications for Voluson i / Voluson e" on page 2-2.
  • Page 91: Dock Cart / Dock Cart Light / Voluson Station Voltage Configuration

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-4-4 Dock Cart / Dock Cart light / Voluson Station Voltage Configuration NOTE: Only required when using the Voluson i / Voluson e with the optional Dock Cart, Dock Cart light or Voluson Station.
  • Page 92: Emi Protection

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-4-5 EMI Protection The Voluson i / Voluson e Ultrasound Unit has been designed to minimize the effects of Electo-Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system from image artifacts caused by this interference.
  • Page 93 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 3-5 Connection of Auxiliary Devices Table 3-9 below outlines the Voluson i/e hardware installation procedures described in the sub-sections. Table 3-9 Connection Procedures Sub-section Description Page Number Connection and Usage of the optional Modo Cart 3-25 Attaching the Voluson i / Voluson e to the Modo Cart 3-25...
  • Page 94: Connection Of Auxiliary Devices

    WARNING WARNING DO NOT attempt to use different peripherals and accessories other than approved and provided by GE! !! WARNING: Any unauthorized peripherals may cause system failure or damage! After each installation, the leakage currents have to be measured according to...
  • Page 95: Connection And Usage Of The Optional Modo Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-1 Connection and Usage of the optional Modo Cart DO NOT CONNECT any devices to any AC power other than the on-board outlets from WARNING WARNING the transformer, as this might increase leakage currents and sacrifice patient safety. !! WARNING: WARNING WARNING...
  • Page 96 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Attaching the Voluson i / Voluson e to the Modo Cart (cont’d) 3-5-1-1 3.) Connect the L-shaped circular power cable plug of the AC Adapter to the power inlet plug DC IN on the rear of the Voluson i / Voluson e unit (see: Figure 3-20 below).
  • Page 97: Mounting Small Peripherals (E.g., Black & White Printer)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-1-2 Mounting small Peripherals (e.g., Black & White Printer) 1.) Place the B&W Printer on the lower peripheral shelf of the Modo Cart. 2.) Take a Velcro tape and pull it under the shelf (left image at Figure 3-22 below).
  • Page 98: Mounting Large Peripherals (E.g., Color Printer)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-1-3 Mounting large Peripherals (e.g., Color Printer) 1.) Place the Color Printer, VCR or DVD recorder on the middle peripheral shelf of the Modo Cart. 2.) Take a Velcro tape and pull it under the shelf (left image at Figure 3-25 below).
  • Page 99: Connection And Usage Of The Optional Dock Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-2 Connection and Usage of the optional Dock Cart DO NOT CONNECT any devices to any AC power other than the on-board outlets from WARNING WARNING the transformer, as this might increase leakage currents and sacrifice patient safety. !! WARNING: WARNING WARNING...
  • Page 100 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Attaching the Voluson i / Voluson e to the Dock Cart (cont’d) 3-5-2-1 6.) Plug the power cable and the USB-cable into the DVD-drive; the other end of the USB cable plug to an USB port on the Voluson i / Voluson e.
  • Page 101: Installation Of The Probe Mux Box

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-2-2 Installation of the Probe Mux Box 1.) Remove the Probe Holder or the Top Probe Holder. Figure 3-30 Remove the Probe Holder / Top Probe Holder 2.) Loosen (don’t remove) the 2 screws (bottom right side of the top shelf) that hold the RIC Probe Holder and slide the holder as far out as you can to the right.
  • Page 102: Installation Of The Probe Shelf

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-2-3 Installation of the Probe Shelf 1.) Place the Probe shelf and fasten it with the 4 Allen screws. Figure 3-32 Fasten the Probe shelf 3-5-2-4 Installation of the Top Probe Holder 1.) Put the Top Probe Holder into the Top Probe Holder Adaptation.
  • Page 103: Mounting Small Peripherals (E.g., Black & White Printer)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-2-5 Mounting small Peripherals (e.g., Black & White Printer) 1.) Place the B&W Printer on the upper peripheral shelf of the Dock Cart. 2.) Take a Velcro tape and pull it under the shelf. 3.) Pull up the Velcro tape on the left side.
  • Page 104: Mounting Large Peripherals (E.g., Color Printer)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-2-6 Mounting large Peripherals (e.g., Color Printer) 1.) Place the Color Printer or VCR recorder on the lower peripheral shelf of the Dock Cart. 2.) Take a Velcro tape and pull it under the shelf. 3.) Pull up the Velcro tape on the left side.
  • Page 105: Ge Voluson I

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-3 Connection and Usage of the optional Voluson Station DO NOT CONNECT any devices to any AC power other than the on-board outlets from WARNING WARNING the transformer, as this might increase leakage currents and sacrifice patient safety. !! WARNING: WARNING WARNING...
  • Page 106 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Attaching the Voluson i / Voluson e to the Voluson Station (cont’d) 3-5-3-1 If applicable, connect the Probe Mux Box. 3.) Push the probe connector into the probe socket of the Voluson i / Voluson e. 4.) Lift the system’s locking lever upwards to lock the probe connector.
  • Page 107: Installation Gel Condom Holder / Probe Holder For Endocavity Probe

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Attaching the Voluson i / Voluson e to the Voluson Station (cont’d) 3-5-3-1 8.) Press the Circuit breaker on the Voluson Station. To start up the system, press the On/Off button (see: Figure 3-87) on the control panel of the Voluson i / Voluson e system (for ~3 seconds).
  • Page 108: Installation Of The Additional Shelf Peripherals

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-3-3 Installation of the Additional Shelf Peripherals 1.) Place the Additional Shelf and fasten it with the 2 screws. Figure 3-42 Installation Additional Shelf Peripherals 3-5-3-4 Installation of the Console Bottom Shelf 1.) Place the Console Bottom Shelf and fasten it with the 2 screws.
  • Page 109: Installation Of The Patient Monitor

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-3-6 Installation of the Patient Monitor 1.) Screw out 2 screws on the Back Handle. Figure 3-45 screw out 2 screws 2.) Insert the right and the left Adapter and fix it with the 2 screws (see Figure 3-45, above).
  • Page 110: Installation Of The Probe Mux Box Voluson Station

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-3-7 Installation of the Probe Mux Box Voluson Station 1.) Fasten the Probe Mux Box with the 4 screws. Figure 3-48 Installation Probe Mux Box 2.) Push the probe connector into the probe socket of the Voluson i / Voluson e and lift the system’s locking lever upwards to lock the probe connector.
  • Page 111: Mounting Small Peripherals (E.g., Black & White Printer)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-3-8 Mounting small Peripherals (e.g., Black & White Printer) 1.) Screw out the screws of the leveling feet from the printer. 2.) Mount the adapter board with the 4 screws on the printer (leveling feet are also mounted). Figure 3-49 screw of the 4 screws / mount the adapter board NOTE: Pay attention to the right mounting position of the printer.
  • Page 112: Mounting Large Peripherals (E.g., Vcr)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-3-9 Mounting large Peripherals (e.g., VCR) 1.) Screw out the screws of the leveling feet from the VCR. 2.) Mount the adapter board with the 4 screws on the VCR (leveling feet are also mounted). Figure 3-52 mount the adapter board NOTE: Pay attention to the right mounting position of the printer.
  • Page 113: Connecting The Wireless Network Adapter/Interface

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-4 Connecting the Wireless Network Adapter/Interface 3-5-4-1 Connection directly to the Voluson i / Voluson e system The Wireless Network adapter should be connected to the PCMCIA slot at the rear of the Voluson i / Voluson e.
  • Page 114: General Remarks And Hints When Using External Usb-Devices

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-5 General Remarks and Hints when using external USB-Devices Do not connect or disconnect any external USB-devices to or from the system while WARNING WARNING scanning a patient! The appearing dialogs could distract you from the scan! 3-5-5-1 External USB-Devices - Connection When an external USB-storage device (such as a USB-memory stick or an external hard disk) is...
  • Page 115: Connection Directly To The Voluson I / Voluson E System

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-6 Connecting the USB Hub 3-5-6-1 Connection directly to the Voluson i / Voluson e system The USB Hub can be connected to the USB ports on the rear connector panel. NOTICE: Auxiliary equipment with direct mains connection requires galvanic separation e.g.
  • Page 116: Connecting The Dvd+R/Rw Drive

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-7 Connecting the DVD+R/RW Drive 3-5-7-1 Connection directly to the Voluson i / Voluson e system The DVD+R/RW Drive can be connected to the USB ports on the rear connector panel. NOTICE: Auxiliary equipment with direct mains connection requires galvanic separation e.g.
  • Page 117: Connection Via Usb Hub

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Connection directly to the Voluson i / Voluson e system (cont’d) 3-5-7-1 NOTE: The power switch on the rear of the DVD drive (Addonics only) must be set in the External position, as shown below: Internal Power OFF...
  • Page 118: Voluson I / Voluson E Mounted On Modo Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-7-5-1 After physical connection 1.) Press the USB button on the keyboard. The “Connect USB and Network Drives” dialog window appears (Figure 3-61 below). Figure 3-61 DVD drive “not used” 2.) If the [Connect to] field shows not used (see: Figure 3-61, above), double-click the “DVD line”.
  • Page 119: Voluson I / Voluson E Mounted On Dock Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-8 Connecting the USB Flash Memory Stick 3-5-8-1 Connection directly to the Voluson i / Voluson e system The USB Flash Memory Stick may be connected to the USB ports on the rear connectors panel. Figure 3-63 Connection Scheme - USB Flash Memory Stick An external USB Flash Memory Stick can be connected once the system is powered ON, or after shutdown.
  • Page 120: Voluson I / Voluson E Mounted On Dock Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-9 Connecting the external USB Hard disk (Handydrive) 3-5-9-1 Connection directly to the Voluson i / Voluson e system The external “Handydrive” HDD may be connected to the USB ports on the rear connectors panel. USB cable Figure 3-64 Connection Scheme - external USB HDD (Handydrive) An external USB Hard disk can be connected once the system is powered ON, or after shutdown.
  • Page 121: Connection Via Usb Hub

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-10 Connecting the Black & White Printer 3-5-10-1 Connection directly to the Voluson i / Voluson e system The B/W Printer can be connected to the USB ports (upper or lower) on the rear connectors panel. power socket USB cable...
  • Page 122: Voluson I / Voluson E Mounted On Modo Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-11 Connecting the Color Printer 3-5-11-1 Connection directly to the Voluson i / Voluson e system The Color Printer can be connected to the USB ports (upper or lower) on the rear connectors panel. power socket USB cable...
  • Page 123: Connecting The Deskjet Color Printer

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-12 Connecting the DeskJet Color Printer 3-5-12-1 Connection via Bluetooth Adapter The DeskJet Color Printer can be connected to an external, non-isolated power source. The bluetooth adapter should be directly connected to the upper USB port on the Voluson i / Voluson e rear connectors panel.
  • Page 124 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Connection via Bluetooth Adapter (cont’d) 3-5-12-1 NOTE: DO NOT connect the Bluetooth adapter via USB hub, to the Modo Cart or to the Dock Cart or to the Voluson Station. This is not an approved configuration.
  • Page 125: Direct Connection To The Voluson I / Voluson E System

    3-5-13 Connecting the VCR (VideoCassette Recorder) NOTE: There are 2 types of Mitsubishi HS-MD3000 VCR (PAL and NTSC) approved by GE. 3-5-13-1 Direct Connection to the Voluson i / Voluson e system Use the standard VCR type (PAL or NTSC) that is suited to your region.
  • Page 126: Voluson I / Voluson E Mounted On Modo Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Direct Connection to the Voluson i / Voluson e system (cont’d) 3-5-13-1 Isolation Transformer power socket Power Supply Video Converter (rear) Video Converter (front) Voluson i/e (rear) Figure 3-70 Connection Scheme - VCR with Video Scan Converter Ophit When connecting the VCR/Video Converter directly to the system, it is necessary to use an additional CAUTION power source.
  • Page 127 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Direct Connection to the Voluson i / Voluson e system (cont’d) 3-5-13-1 NOTE: Dip switches are located on the front respectively beneath the video converter, see Figure 3-71 below. Video Scan Converter Canopus Video Scan Converter Ophit Input Select button...
  • Page 128: Voluson I / Voluson E Mounted On Dock Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-13-2 Voluson i / Voluson e mounted on Modo Cart NOTE: For instructions see: Section 3-5-1-3 "Mounting large Peripherals (e.g., Color Printer)" on page 3-28. The VCR must be powered from the Modo Cart itself, via one of the isolated peripheral power cables provided on the Modo Cart.
  • Page 129: Voluson I / Voluson E Mounted On Voluson Station

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-13-4 Voluson i / Voluson e mounted on Voluson Station NOTE: For instructions see: Section 3-5-3-9 "Mounting large Peripherals (e.g., VCR)" on page 3-42. Power Socket from Voluson Station Figure 3-73 Connection Scheme The VCR must be powered from the Voluson Station itself, via one of the isolated peripheral power cables provided on the Voluson Station.
  • Page 130: Connecting The Dvd Recorder

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-14 Connecting the DVD Recorder The Sony DVO-1000MD DVD recorder is only able to read and write on DVD+RW (ReWriteable) NOTICE media! Please check on your DVD case before using. !! NOTICE: 3-5-14-1 Direct Connection to the Voluson i / Voluson e system...
  • Page 131: Voluson I / Voluson E Mounted On Modo Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Direct Connection to the Voluson i / Voluson e system (cont’d) 3-5-14-1 power socket isolation transformer Power Supply Video Converter (rear) Video Converter (front) Voluson i/e (rear) Figure 3-75 Connection Scheme - DVD Recorder with Video Scan Converter Ophit When connecting the DVD Recorder/Video Converter directly to the system, it is necessary to use an CAUTION additional power source.
  • Page 132 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Direct Connection to the Voluson i / Voluson e system (cont’d) 3-5-14-1 NOTE: Dip switches are located beneath the video converter, as shown in Figure 3-76 below. Video Scan Converter Canopus Video Scan Converter Ophit Input Select button Dip Switches...
  • Page 133: Voluson I / Voluson E Mounted On Dock Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-14-2 Voluson i / Voluson e mounted on Modo Cart NOTE: For instructions see: Section 3-5-1-3 "Mounting large Peripherals (e.g., Color Printer)" on page 3-28. The DVD recorder must be powered from the Modo Cart itself, via one of the isolated peripheral power cables provided on the Modo Cart.
  • Page 134 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-14-4 Voluson i / Voluson e mounted on Voluson Station NOTE: For instructions see: Section 3-5-3-9 "Mounting large Peripherals (e.g., VCR)" on page 3-42. Power Socket from Voluson Station Figure 3-78 Connection Scheme The DVD recorder must be powered from the Voluson Station itself, via one of the isolated peripheral power cables provided on the Voluson Station.
  • Page 135: Adjustment Of The Dvd Recorder Settings

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-14-5 Adjustment of the DVD Recorder Settings 3-5-14-5-1 Changing TV-System (Video format) Table 3-10, “TV-System and Region Code,” on page 3-67. NOTE: To check Video format needed, see: 1.) If not already done, switch on the Sony DVO-1000MD DVD recorder. 2.) Press the MENU button (A) located on the front of the DVD recorder, see Figure 3-79 below.
  • Page 136 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Adjustment of the DVD Recorder Settings (cont’d) 3-5-14-5 6.) Press the ↓ (down) button repeatedly, until [MENU GRADE] is highlighted. Figure 3-81 “Setup Menu” - [MENU GRADE] highlighted 7.) Press the → (right) button and select [ENHANCED]. 8.) Press the ←...
  • Page 137: Change The Region Code

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-5-14-5-2 Change the Region Code Table 3-10, “TV-System and Region Code,” on page 3-67. NOTE: To specify the correct region code, see: 1.) If not already done, switch on the Sony DVO-1000MD DVD recorder. 2.) Press the MENU button (A) located on the front of the recorder, see Figure 3-79 on page 3-65.
  • Page 138: Completing The Set Up

    Voluson i / Voluson e System User Training. Use only the power cords, cables and plugs provided by or designated by GE to connect the unit to the power source. Prior to connect the Voluson i / Voluson e unit to a power source, verify compliance with all CAUTION electrical and safety requirements.
  • Page 139: Power On / Boot Up

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Voltage Level Checks (cont’d) 3-6-1-1 The following voltage level checks are required whenever the Voluson i / Voluson e system is used with the Dock Cart or the Voluson Station: 1.) Check the Fuses and Voltage Output Selector setting located on the power supply of the Cart.
  • Page 140 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-6-2 Power On / Boot Up (cont’d) NOTE: Through the protection plate near the plug, the power cable can only be plugged in one way. protection plate L-shaped circular plug of main power cable connected to DC IN Figure 3-86 main power cable connected to Voluson i / Voluson e...
  • Page 141: During A Normal Boot, You May Observe

    Software 6.x.x and 7.x.x with Software 8.x.x and higher Figure 3-89 GE wallpaper with progress bar 5.) The software initiates and sets up the Front-End electronics and the rest of the scanner. 6.) The Keyboard backlight is lit.
  • Page 142: Power Off/ Shutdown

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-6-3 Power Off/ Shutdown • Standby Mode NOTE: The Standby Mode on the Voluson i / Voluson e is available with Software 7.x.x. and higher. 1.) If not already in read mode, freeze the image. 2.) Press the On/Off button (see: Figure 3-87) on the control panel.
  • Page 143: Internal Lithium Ion Battery

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-6-4 Internal Lithium Ion Battery One battery is standard delivered with the Voluson i / Voluson e. This lithium ion battery provides power when an AC power source is not available. You can expect one hour of battery life with a fully charged battery.
  • Page 144: Charging The Battery

    NOTE: Lithium ion batteries last longer than conventional batteries and do not require replacement as often. Only use the batteries designed to work with the Voluson i / Voluson e, and authorized by GE Medical Systems. To avoid the battery bursting, igniting, or fumes from the battery causing equipment damage,...
  • Page 145: External Battery Charger

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-6-4-1-2 External Battery Charger NOTE: is a stand-alone desktop smart-battery charger with the added ability to recalibrate the CH5000 fuel gauge when detected to be out of calibration. The charger is specifically designed to operate with the Voluson i / Voluson e smart-battery.
  • Page 146: Recalibrating The Battery

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 3-12 Battery Charger LED Indicators and Status LED Indicator Status Green Flashing Battery charging Green Solid Battery fully charged Blue flashing Battery in calibration mode Blue Solid Battery fuel gauge calibrated Red Flashing Battery fuel gauge in need of recalibration Red Solid...
  • Page 147 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-6-4-2-1 Recalibration by means of Internal Battery Charger 1.) Press the UTILITIES key and select the SYSTEM from the menu area on the left side of the screen. 2.) Select the page on the screen.
  • Page 148: Recalibration By Means Of External Battery Charger

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-6-4-2-2 Recalibration by means of External Battery Charger NOTE: The LEDs in the Chargers status window will provide status information as detailed in Table 3-11 on page 3-73. If the battery is in need of “fuel gauge” recalibration, the red LED will flash upon insertion of the battery into the battery bay.
  • Page 149: Transducer Connection

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-6-5 Transducer Connection When the probe is connected, it is automatically activated. Once connected, the probes can be selected for different applications. 1.) Inspect the probe and probe socket to verify that it is free of debris. 2.) Press the probe connector locking lever downwards.
  • Page 150: Printer Installation

    3-91. The Bluetooth Printer Connection set as well as the Color Deskjet printer MUST NOT be installed CAUTION by the user! For installation please contact your local distributor or GE service representative. !! CAUTION: 3-80 Section 3-7 - Printer Installation...
  • Page 151: Installing Digital Black & White Printer Sony Up-D897

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-7-1 Installing Digital Black & White Printer Sony UP-D897 1.) Power off/Shutdown the system as described in: Section 3-6-3 on page 3-72. 2.) Connect the printer as described on page 3-51.
  • Page 152 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Install the UP-D897 printer software/driver (cont’d) 3-7-1-1 3-98, click on the BROWSE button to NOTE: If path (R:\system\..) is different from what is shown in Figure search correct driver. R:\system\periph\SonyUP-D897 verify correct path, otherwise [Browse] for R:\system\periph\SonyUP-D897...
  • Page 153: Installing Digital Color Printer Sony Up-D23Md Or Up-D25Md

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-7-2 Installing Digital Color Printer Sony UP-D23MD or UP-D25MD 1.) Power off/Shutdown the system as described in: Section 3-6-3 on page 3-72. 2.) Connect the printer as described on page 3-52.
  • Page 154 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Install the UP-D23MD/UP-D25MD printer software/driver (cont’d) 3-7-2-1 3-98, click on the BROWSE button to NOTE: If path (R:\system\..) is different from what is shown in Figure search correct driver. R:\system\periph\SonyUP-D23MD verify correct path, otherwise [Browse] for R:\system\periph\SonyUP-D23MD...
  • Page 155: Installing Mitsubishi B&W Printer P95 And Color Printer Cp30

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-7-3 Installing Mitsubishi B&W Printer P95 and Color Printer CP30 NOTE: "Mouse functions" can be performed by using the trackball for moving the cursor. "Normal select" (Click) = left/right trackball key; "Opening a context menu" = upper trackball key NOTE: To use the Mitsubishi Printer with older Systems and Software versions a printer driver has to be installed.
  • Page 156 "C:\Utilities\PrinterDriver\Win2000_XP" for the Color Printer. Path depeD95 or F:\Win2000_XP. (where F: stands for CD/DVD Drive) Then click on NEXT. NOTE: If the "Printer Driver" Disk is defective or not available, download driver from GE folders http://libraries.ge.com/foldersIndex.do?entity_id=15869258101&sid=101&SF=1#15869248101 and extract files to CD or USB disk.
  • Page 157: Printer Installation Manually

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-7-4 Printer Installation manually 1.) Press the UTILITIES key on the keyboard. 2.) In the “Utilities” menu on the left side of the screen, select SYSTEM to invoke the setup desktop. 3.) Select the page.
  • Page 158 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-7-4 Printer Installation manually (cont’d) 7.) Select the ‘Local Printer‘, deselect “Automatically install Plug and Play printer” and then click NEXT. Figure 3-108 Add Printer Wizard 8.) Select the corresponding Printer Port (e.g., Figure 3-109 = USB001) and click NEXT.
  • Page 159 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-7-4 Printer Installation manually (cont’d) 10.)Use the BROWSE button to search the Printer Driver path (R:\system\periph\xxxx). 11.)Click OPEN. Select the “xxx.inf” file and click OPEN again. R:\system\periph\SonyUP-D897 Figure 3-110 Select Printer Driver path (R:\system\periph\..) / verify selected Printer Driver path 12.)Verify the selected Printer Driver path and confirm with OK.
  • Page 160: Ge Voluson I

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-7-4 Printer Installation manually (cont’d) 15.)The “Complete the Add Printer Wizard” window appears on the screen. Figure 3-112 Complete manual Printer Installation 16.)Complete the manual Printer Installation with the FINISH button. NOTE: If the message “The software you are installing for this hardware has not passed Windows ..”...
  • Page 161: Adjustment Of Printer Settings

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-7-5 Adjustment of Printer Settings 1.) After system restart, press the UTILITIES key on the alphanumeric keyboard and then select the SYSTEM item from the menu area on the left side of the screen. 2.) Select the page.
  • Page 162: Up-D897 - Printer Settings

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-7-5-1 UP-D897 - Printer Settings 1.) Call up the ‘Printer Preferences’; operation see: Section 3-7-5 "Adjustment of Printer Settings" 2.) Select the page (see: Figure 3-114 below) and select: LAYOUT Paper: 960x1280 Orientation: Portrait...
  • Page 163: Up-D23Md/Up-D25Md - Printer Settings

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-7-5-2 UP-D23MD/UP-D25MD - Printer Settings 1.) Call up the ‘Printer Preferences’; operation see: Section 3-7-5 "Adjustment of Printer Settings" Settings for Paper Size MUST match with the used Paper (large/small) and also the right color ink NOTICE cartridge has to be used.
  • Page 164 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL UP-D23MD/UP-D25MD - Printer Settings (cont’d) 3-7-5-2 3.) Select the page. From the “Color Adjust” pop-up menu select: GRAPHICS a.) Color Balance: Cyan = 0; Magenta = 0; Yellow = 0 b.) Gamma Select: Gamma 1 Figure 3-116 Graphics page (Color Balance + Gamma Select) c.) Color Correction: set Printer Hardware Color Correction...
  • Page 165: Mitsubishi Printer Settings

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-7-5-3 Mitsubishi Printer Settings 3-7-5-3-1 Color Printer CP30 1.) Call up the ‘Printer Preferences’; operation see: Section 3-7-5 "Adjustment of Printer Settings" 2.) Select the PAPER page and select: Paper Size: L (large) Orientation: Landscape (recommended when using large paper size) Figure 3-118 Paper page...
  • Page 166: Bluetooth Deskjet - Printer Settings

    Bluetooth Deskjet - Printer Settings The Bluetooth Connection Set as well as the color printer MUST NOT be installed by the user! CAUTION For installation please contact your local distributor or GE service representative. !! CAUTION: 1.) Call up the ‘Printer Properties’; operation see: Section 3-7-5 "Adjustment of Printer Settings"...
  • Page 167: Printer Remote Control Selection

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-7-6 Printer Remote Control Selection To assign a Printer to the remote keys P1, P2 and/or P3: 1.) Press the UTILITIES key on the keyboard. 2.) In the “Utilities” menu on the left side of the screen select SYSTEM to invoke the setup desktop. 3.) Select the page.
  • Page 168: System Configuration

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 3-8 System Configuration 3-8-1 System Setup Modifications of system parameters are supported by diverse dialog pages and windows on the system setup desktop: • General - Date, Time, Clinic Name, Language, Screen saver, etc. •...
  • Page 169: How To Enter Date And Time

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-8-1-1 How to enter Date and Time Select the page in the System Setup see: Figure 3-122. GENERAL 1.) Select the “Date Format” (only one can be active). 2.) Click the DATE/TIME button to activate a sub dialog window to enter date, time and time zone. 3.) Click the TIME FORMAT button to activate a sub dialog window to choose preferred time format.
  • Page 170: How To Adjust Function Of The Footswitch (Left/Right)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-8-1-4 How to adjust function of the Footswitch (Left/Right) NOTE: The Footswitch is available with Software 7.x.x. and higher Select the page in the System Setup. USER SETTINGS Figure 3-123 System Setup - User Settings 1.) Select desired function of the Footswitch Left and Right.
  • Page 171: How To Activate Screen Lock

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-8-1-5 How to activate Screen Lock Select the page in the System Setup see: Figure 3-122. GENERAL 1.) Check mark “Screen Lock”. a.) If no password previously entered, the following dialog appears. Figure 3-124 Screenlock Password b.) Enter “New Password”.
  • Page 172: Measure Setup

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-8-2 Measure Setup Modifications of system parameters are supported by diverse dialog pages and windows on the measure setup desktop. Parameters and possible adjustments mostly depend on the selected Application! NOTICE !! NOTICE: •...
  • Page 173: Approved Optional Peripherals

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-8-3 Approved Optional Peripherals The following table lists the internal peripherals available for use with the Voluson i / Voluson e portable ultrasound system and details their method of connection. Table 3-13 Approved Peripherals Device...
  • Page 174: External I/O Connectors

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-8-4 External I/O Connectors Figure 3-126 External I/O Connectors Table 3-14 External I/O Connector Descriptions Item Connector Name Table Description two interchangeable USB-2.0 ports Table 3-15 USB 1 and USB 2 (Printers, DVD+R/RW drive and other peripheral) Connector for Voluson Station Docking Adapter Docking connector...
  • Page 175: External I/O Pin Outs

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-8-4-1 External I/O Pin Outs Table 3-15 USB 1, 2 Connectors Pin No Output Signal Description USB Power Supply - Data USB Data (-) + Data USB Data (+) USB Power Ground Table 3-16 VGA OUT Connector, Sub-D 15 Pin...
  • Page 176: External Cables - Maximum Lengths

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-8-5 External Cables - Maximum Lengths Table below shows maximum permitted cable length of external cables, according to IEC60601-1-2. Table 3-18 maximum cable lengths Description Maximum Cable Length Type Probe cable 2.5 m shielded...
  • Page 177: Connectivity Setup

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 3-11 Connectivity Setup The Voluson i / Voluson e portable ultrasound system can be connected to various connectivity devices. The following sections describe how to connect the system to a remote archive/work station or a DICOM®...
  • Page 178: Dataflow Examples

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-11-1-2 Dataflow Examples Voluson i/e The local database is used for patient archiving. Images are stored to internal hard drive. Voluson i/e The local database is used for patient archiving. Images are stored to a DVD/CD.
  • Page 179: Stand-Alone Voluson I / Voluson E

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-11-1-3 Stand-alone Voluson i / Voluson e If digital images or 3D/4D data sets are stored, they should be saved in Sonoview (Image Management System software). For Sonoview - Image Management functionality refer to the Basic User Manual of the Voluson i / Voluson e portable ultrasound system.
  • Page 180: Wireless Network Configuration

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-11-1-7 Wireless Network Configuration The following procedure is used to configure the Voluson i / Voluson e system for a wireless network environment. To configure the Voluson i / Voluson e ultrasound unit to work with the WLAN adapter / NOTICE WLAN Interface, the hospital´s network administrator must provide the required information.
  • Page 181 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Wireless Network Configuration (cont’d) 3-11-1-7 5.) Select the tab. PROFILE MANAGEMENT [New] button [Modify] button Figure 3-130 Profile Management 6.) Click the NEW button to create a new, or the MODIFY button to modify the currently selected profile. NOTE: It is recommended to modify the “Default”...
  • Page 182: Setting General Parameters

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-11-1-7-1 Setting General Parameters The “Profile Management - General” window (see: Figure 3-131 below) enables you to set parameters that prepare the client adapter for use in a wireless network. This window appears after you click NEW or MODIFY on the Profile Management (Figure 3-130 on page...
  • Page 183: Setting Advanced Parameters

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-11-1-7-2 Setting Advanced Parameters The “Profile Management - Advanced” window (see: Figure 3-132 below) enables you to set parameters that control how the client adapter operates within an infrastructure or ad hoc network. To open this window, click the tab from any Profile Management window.
  • Page 184: Setting Security Parameters

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 3-21 Profile Management Advanced Parameters (cont´d) Parameter Description Specifies the frequency and rate at which your client adapter should transmit packets to or receive packets from access points. Default: All options selected Wireless Mode Note: When more than one option is selected, the client adapter attempts to use the wireless modes in...
  • Page 185: Network Profiles

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-11-1-8 Network Profiles 1.) Connect the Wireless Network adapter as described in Section 3-5-4 on page 3-43 or connect the network cable. 2.) Press the UTILITIES key on the keyboard once to display the Utilities menu. 3.) Select SYSTEM and then open the page on the System Setup desktop screen.
  • Page 186: Network Ip Address Configuration

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 3-12 Network IP Address Configuration NOTE: Following Information must be provided by customer or hospital engineer before you can start: A Station name, AE Title, IP address and Port Number for the Voluson i / Voluson e. The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION.
  • Page 187: Dicom /Sonoview Configuration

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-12-2 Dicom /SonoView Configuration ® DICOM® is the abbreviation of Digital Imaging and Commnications in Medicine. This is the industrial standard for communication of images and other information between medical devices on the network. Using the DICOM®...
  • Page 188: Map Network Drive

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 3-12-3 Map Network Drive Press the USB key on the keyboard, the “Connect USB and Network Drives” dialog window (see: Figure 3-56 on page 3-44) is displayed. 1.) Select the MAP NETWORK DRIVE button to open a dialog where the system can be connected to a shared network drive of another server.
  • Page 189: Connectivity Setup Worksheet

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 3-13 Connectivity Setup Worksheet Site System Information Comments: Floor: Site: Dept: Room: Type: Voluson i/eSN: REV: CONTACT INFORMATION Name Title E-Mail Address Phone Remote Archive Setup TCP/IP Settings Name - AE Title: System IP Settings IP Address:...
  • Page 190 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 3-13 Connectivity Setup Worksheet (cont’d) Voluson i / Voluson e IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default ROUTER1 ROUTER2...
  • Page 191: Paperwork

    Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. Chapter 3 - Setup Instructions...
  • Page 192 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 3-122 Section 3-14 - Paperwork...
  • Page 193: Overview

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the Voluson i / Voluson e scanner diagnostics by using the built-in service software, and power supply adjustments. Table 4-1 Contents in Chapter 4 Section...
  • Page 194: System Features

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-3-1 System Features 4-3-1-1 Control Panel Figure 4-1 Control Panel of the Voluson i / Voluson e Menu Control keys (Soft-menu buttons) ON/OFF Switch alphanumeric keyboard with main function keys (e.g., TGC - Slider Controls Utilities, Bodymarks, Power up/down, etc.) Mode keys...
  • Page 195: Menu Control

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-3-1-2 Menu Control Figure 4-2 Menu Control Area - 2D Main Menu 1.) User Program: shows all setting for the active application. The active one is highlighted 2.) Additional functions which are supported by the selected Scan mode. 3.) Sub Menu: to adjust settings of the selected “Main”...
  • Page 196: Monitor Display

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-3-1-3 Monitor Display Figure 4-3 Monitor Display Tour Table 4-2 Monitor Display Features 1. Logo 14. Image Area 2. Patient Name (Last-, First-, Middle Name) 15. Measurement results 3. Patient ID-number ; GA (Gestational Age) 16.
  • Page 197: Functional Checks

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 4-4 Functional Checks For a basic functional check of the system’s different modes, following pages will familiarize you with image optimization for: • 2D Mode (B Mode) • M Mode MCFM Mode •...
  • Page 198: 2D Mode Checks

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-1 2D Mode Checks Figure 4-4 2D Main and 2D Sub Table 4-3 2D Mode Functions Step Task Expected Results Rotate the ACTIVE MODE rotary control to adjust the sensitivity (brightness) 2D Mode Gain of the entire image.
  • Page 199 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 4-3 2D Mode Functions Step Task Expected Results This function allows the adjustment of the Volume O-Axis position of 3D ß-VIEW probes in 2D Mode. The green line in the displayed symbol indicates the position of the acoustic block.
  • Page 200 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 4-3 2D Mode Functions Step Task Expected Results Dynamic Range controls how echo intensities are converted to shades of DYN CONTR gray, thereby increasing the adjustable range of contrast. The signals of the neighboring pulses are less weighted for the display of the LINE FILTER actual pulse which considerably improves the detail lateral resolution and...
  • Page 201: M Mode Checks

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-2 M Mode Checks Figure 4-5 M Main and M Sub Table 4-4 M Mode Functions Step Task Expected Results Cursor Position Adjust the M Cursor position with the TRACKBALL in the 2D Single image. Activation of M Mode Press the right or left trackball key to activate both Modes (2D/M).
  • Page 202: Mcfm Mode Check

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-2-1 MCFM Mode Check By pressing the M control and the CFM control, the MCFM mode is switched on in preparation mode. The M-cursor with M-Color window appears in the active 2D image. Figure 4-6 MCFM Main and MCFM Sub Menu For further details refer to the Voluson i / Voluson e Basic User Manual.
  • Page 203: Spectral Doppler Mode Checks

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-3 Spectral Doppler Mode Checks Figure 4-7 PW Main and PW Sub Table 4-5 Spectral Doppler Mode (PW) Functions Step Task Expected Results Adjust the G-POS (Gate position) or the ANGLE (cursor can be turned in both Gate Position and Angle directions without stop) with the trackball.
  • Page 204 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 4-5 Spectral Doppler Mode (PW) Functions Step Task Expected Results The Velocity Range display is governed by the pulse repetition frequency (PRF): PRF (Pulse Repetition Frequency) Press the Soft-menu buttons to adjust the velocity range desired. FREQUENCY Serves for selection of the required transmit frequency for actual gate position.
  • Page 205: Color Flow Mode Checks

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-4 Color Flow Mode Checks NOTE: Different menus are displayed depending on which Color Flow Mode (Color or Power) is selected. Figure 4-8 CFM Main and Color Sub Menu Table 4-6 Color Flow Mode (Color or Power Doppler) Functions HD Step...
  • Page 206 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 4-6 Color Flow Mode (Color or Power Doppler) Functions HD Step Task Expected Results The baseline shift can be used to prevent aliasing in one flow direction similar to the Doppler BASELINE baseline shift.
  • Page 207: Volume Mode Checks

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-5 Volume Mode Checks Static 3D, 4D Real Time,TUI*, VCI*, RT 4D Biopsy*, SonoAVC, SonoVCAD heart*, NOTICE SonoVCAD labor and STIC* are optional. !! NOTICE: If these options are not a part of the system configuration, the appendant checks can be obmitted. The options “RT 4D Biopsy”...
  • Page 208 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Pre-Volume Mode Functions (cont’d) 4-4-5-1 Table 4-7 Pre-Volume Mode Functions Step Task Expected Results 3D RENDERING 3D Static volume acquisition + rendered 3D image (also in combination Color or PD Mode) SECTIONAL PLANES 3D Static volume acquisition resp.
  • Page 209: Functions After The 3D Acquisition

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-5-2 Functions after the 3D Acquisition Figure 4-10 3D/4D Main - Sectional Planes and Image Rendering Figure 4-11 SonoAVC / SonoVCAD Table 4-8 Functions after the 3D Acquisition Step Task Expected Results If not currently active, press the MENU POINTER [Cursor] key near the trackball.
  • Page 210 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 4-8 Functions after the 3D Acquisition Step Task Expected Results Rotation about Z-axis If not currently active, press the MENU POINTER [Cursor] key near the trackball. Now move the cursor to this icon on the screen and then press SET (right or left trackball key).
  • Page 211: Sub Menu

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-5-3 Sub Menu Figure 4-12 3D/4D Sub Menu Table 4-9 Sub Menus Step Task Expected Results PROBE ORIENTATION Adjust and activate the display of direction markers at border of the image. Render PARAMETER To select the Render Mode (Basic Mode and Render Algorithm).
  • Page 212: Using Cine

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-6 Using Cine 4-4-6-1 Activating Cine Press FREEZE, then roll the TRACKBALL to display the images of the stored sequence one by one. 4-4-6-2 Cine-Split Function (Multiple Format) After FREEZE of a sequence in 2D Mode two or four different images of the sequence can be displayed simultaneously in Dual or Quad Display Mode respectively.
  • Page 213: Generic Measurements

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-7 Generic Measurements NOTE: Different menus are displayed depending on: • the currently selected Application and Display Mode, • the selected “Study”, MEASURE & CALC • and the settings in the Measure Setup - page.
  • Page 214: Circumference/Area Measurements

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-7 Generic Measurements (cont’d) 4-4-7-2 Circumference/Area Measurements 1.) Press the CALIPER key and then select the GENERIC AREA in the menu area. 2.) Select the corresponding item from the menu area. An active cursor displays. 3.) To position the active cursor, move the TRACKBALL.
  • Page 215: Measurements In Spectral Doppler Mode

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-7 Generic Measurements (cont’d) 4-4-7-4 Measurements in Spectral Doppler Mode NOTE: The Spectral Doppler image is displayed based on time (X-axis) and velocity (Y-axis). 4-4-7-4-1 Auto Trace 1.) Press the CALIPER key. 2.) Select AUTO TRACE item from the menu area on the left side of the screen.
  • Page 216: Calculations

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-8 Calculations The Voluson i / Voluson e system supports calculation packages and application-oriented patient Worksheets (Reports) for following applications: • Abdomen Calculations • Small Parts Calculations • Obstetric Calculations •...
  • Page 217: Probe/Connectors Usage

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-9 Probe/Connectors Usage 4-4-9-1 Connecting a probe 1.) Place the probe's carrying case on a stable surface and open the case. 2.) Carefully remove the probe and unwrap the probe cable. 3.) Press the probe connector locking lever downwards.
  • Page 218: Image Management (Sonoview)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-10 Image Management (Sonoview) For Sonoview - Image Management functionality refer to the Basic User Manual of Voluson i/e. It talks about several topics: • Clipboard • Sending Exams •...
  • Page 219: Formatting Media

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-4-11 Formatting Media To erase the backup media DVD/CD+RW, press the ARCHIVE key on the Control panel. The Sonoview screen appears on the monitor; see: Figure 4-13 on page 4-26.
  • Page 220: Backup And Restore Database, Preset Configurations And Images

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 4-5 Backup and Restore Database, Preset Configurations and Images 1.) Press the UTILITIES key on the keyboard. The menu area changes to the Utilities menu. 2.) Select the SYSTEM item from the menu area to activate the setup desktop screen. 3.) Select the page.
  • Page 221: Save User Settings Only (Application Settings)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-5-1 Save User Settings Only (Application Settings) The User Settings contains: • Application Settings • User Programs • Auto Text • 3D/4D Programs 1.) Insert a DVD/CD+(R)W into the drive. 2.) Press the UTILITIES key on the keyboard.
  • Page 222: Load User Settings Only (Application Settings)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-5-2 Load User Settings Only (Application Settings) The loading procedure overwrites the existing application settings on the local hard drive. Make sure to CAUTION insert the correct DVD/CD. Additionally you can load the backup from D:\User Settings. !! CAUTION: 4-5-2-1 Preparations...
  • Page 223: Load "Complete Backup

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-5-2 Load User Settings Only (Application Settings) (cont’d) 7.) Select the desired loading procedure: Load “Complete Backup” on page 4-31 Load only parts of the “Complete Backup” on page 4-31 4-5-2-2 Load “Complete Backup”...
  • Page 224: Save Full Backup (Presets, Configurations & Application Settings)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-5-3 Save Full Backup (Presets, Configurations & Application Settings) A full backup always contains the following data • Patient demographic and exam data (database containing the patient data and measurements) •...
  • Page 225 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-5-3 Save Full Backup (Presets, Configurations & Application Settings) (cont’d) When the Full Backup is stored on a network drive (to map a network drive see: Section 3-12-3 "Map Network Drive"...
  • Page 226: Load Full Backup (Presets, Configurations & Application Settings)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-5-4 Load Full Backup (Presets, Configurations & Application Settings) It is recommended to backup data before an upgrade; see: Section 4-5-3 on page 4-32. WARNING WARNING The “Full Backup” loading procedure replaces (overwrites) the existing data on the !! WARNING: local hard drive of the Voluson i / Voluson e system!
  • Page 227 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-5-4 Load Full Backup (Presets, Configurations & Application Settings) (cont’d) 1.) Press the UTILITIES key on the keyboard. The menu area changes to the Utilities menu. 2.) Select the SYSTEM item from the menu area to activate the setup desktop screen. 3.) Select the page and click the LOAD button of the “Full Backup”...
  • Page 228: Delete Full Backup (Presets, Configurations & Application Settings)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-5-4 Load Full Backup (Presets, Configurations & Application Settings) (cont’d) 8.) Click the NEXT button and then select YES to start, or NO to cancel the restore procedure. Figure 4-28 Start Restore Backup now? WARNING WARNING When clicking “YES”, the current data on the system will be permanently replaced by...
  • Page 229: Archiving Images

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 4-5-6 Archiving Images 1.) Press the ARCHIVE key on the control panel to enter Sonoview. 2.) Insert the DVD/CD+(R)W into drive. If required, format/erase the media, see Section 4-4-11 "Formatting Media" on page 4-27.
  • Page 230: Software Configuration Checks

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 4-6 Software Configuration Checks 4-6-1 System Setup Press the UTILITIES key and select the SYSTEM item from the “Utilities” menu - which is shown in the menu area on the left side of the screen. The System Setup desktop offers different pages to check: Table 4-10 Software Configuration (System Setup) Checks Step...
  • Page 231: Peripheral Checks

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 4-7 Peripheral Checks Check that peripherals work as described below: Table 4-12 Peripheral Checks Step Task to do Expected Result(s) Press the FREEZE key. Stop image acquisition. The image displayed on the screen is printed, saved or Press the P1, P2 and/or P3 key on the Control Panel.
  • Page 232: Site Log

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 4-8 Site Log Table 4-13 Voluson i / Voluson e - Site Log (Paper Documentation) Date Service person Problem Comments 4-40 Section 4-8 - Site Log...
  • Page 233: Overview

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains Voluson i / Voluson e’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution System (PDS) and probes. Table 5-1 Contents in Chapter 5 Section Description...
  • Page 234: General Information

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 5-2 General Information The Voluson i / Voluson e is a compact, fully portable, digital beamforming curved and linear array ultrasound imaging scanner. Weighing only 5 kg / 11 lbs. (without peripherals), this laptop-style system is extremely versatile.
  • Page 235 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 5-2 General Information (cont’d) Among other significant features of the Voluson i / Voluson e portable ultrasound system are the following: Integrated, low-power Front End (uses advanced ASIC and FGPA technologies). Open connectivity using dual PCMCIA interface and USB Bluetooth - wireless connectivity Low power (less than 8W), high performance 15"...
  • Page 236: Ge Voluson I

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 5-2 General Information (cont’d) The Voluson i / Voluson e uses digital beamforming technology which provides high resolution and high penetration performance. It is a general purpose, software controlled diagnostic portable ultrasound scanner.
  • Page 237: Description Of Voluson I / Voluson E Operating Modes

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-1 Description of Voluson i / Voluson e Operating Modes 5-2-1-1 B-Mode or 2D-Mode B-Mode or 2D-mode is a two-dimensional image of the amplitude of the echo signal. It is used for location and measurement of anatomical structures and for spatial orientation during operation of other modes.
  • Page 238: Color Doppler Mode

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-1-3 Color Doppler Mode Color Doppler is used to detect motion presented as a two-dimensional display. There are two applications of this technique: • Color Flow Mode (C) - used to visualize blood flow velocity and direction •...
  • Page 239: 4D Imaging

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-2 3D/4D Imaging The Voluson i / Voluson e Ultrasound System will be used to acquire multiple, sequential 2D images which can be combined to reconstruct a three dimensional image. These 3D images are useful in visualizing three-dimensional structures, and in understanding the spatial or temporal relationships between the images in the 2D sequence.
  • Page 240: Block Diagram Voluson I / Voluson E

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-3 Block diagram Voluson i / Voluson e Microphone not available with SW 8.1.2 and higher Figure 5-3 Top Level Architecture Section 5-2 - General Information...
  • Page 241: Dataflow Control Description

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-4 Dataflow Control Description This section describes functions of Voluson i / Voluson e Boards vs different Operation Modes. • RBF - Transmitter/Receiver “Beamformer” Board • RBI - Radio Frequency “Controller” Board 5-2-4-1 B-Mode 1.) RBI...
  • Page 242: Special B Mode Techniques

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-4-1-1 Special B Mode Techniques a.) HI (Coded Harmonic Imaging) In one method of HI the RX-Frequency is doubled, so that the radial resolution is increased due to the higher RX-Frequency. The second method of HI is pulse-inversion: 2 TX-Beams are shot to the same Tissue-location, one with positive, one with negative polarity.
  • Page 243: D-Mode Autotrace (Draws Pc-Calculated Envelope To D-Spectrum)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-4-4 D-Mode Autotrace (draws PC-calculated envelope to D-Spectrum) 1.) RBI PRF-generator; see: 5-2-4-1 B-Mode After DC-cancelling the signal is mixed with RX-Frequency and brought to LF-Spectrum, where the LowPassFilter cuts HF. Mixer and Magnitude-Calculator arrange Complex Demodulation.
  • Page 244: Description Of Software Options

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-5 Description of Software Options To activate software options: 1.) Press the UTILITIES key on the keyboard. The menu area changes to the Utilities menu. 2.) Select the SYSTEM item from the menu area and then select the page.
  • Page 245 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-5 Description of Software Options (cont’d) Table 5-3 Software Options at Voluson e SW-Option at Voluson e Description ® ® DICOM® DICOM 3 (Interface for DICOM 3 Standard) Real Time 4D 4D Basic (4D rendering only!) (4D rendering only, no 3D Mode and no access to volume data) 3D Mode...
  • Page 246: Mode

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-5-1 3D Mode refer to: Section 5-2-2 "3D/4D Imaging" on page 5-7 5-2-5-2 Real Time 4D Real Time 4D mode is obtained through continuous volume acquisition and parallel calculation of 3D rendered images.
  • Page 247: Dicom

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-5-5 DICOM® Voluson i / Voluson e software package providing following DICOM® functionality: • Storage Service Class • Print Management Service Class • Structured Reporting Service Class • Storage Commit Service Class •...
  • Page 248: Sonoavc - Sono Automated Volume Count

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-5-9 SonoAVC - Sono Automated Volume Count This Feature can automatically detect low echogenic objects (e.g., follicles) in a volume of an organ (e.g., ovary) and analyze their shape and volume. From the calculated volume an average diameter can be calculated.
  • Page 249: Sonont

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-2-5-13 SonoNT SonoNT is an additional function for manual NT (Nuchal Translucency) measurement. This function supports the user to find the correct position for the NT measurement. The user can switch between NT Method “Manual” and “Sono NT” (semi-automatic). A box has to be placed for the NT-ROI.
  • Page 250: Front End Processor

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 5-3 Front End Processor 5-3-1 General Information Microphone not available with SW 8.1.2 and higher Bottom Assembly Back End Front End 15‘‘ LCD + RBF (1) Microphone 32 x (Tx+Rx) Baseboard GPM (RPM) RBF (2)
  • Page 251: Radio Frequency Interface (Rbi) Board

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-3-2 Radio Frequency Interface (RBI) Board The Radio Frequency Interface (RBI) Board combines the following sub-boards and functionalities: • FEC (Front End Controller) • RFT (Radio Frequency & Tissue) processor •...
  • Page 252: Probe And Mux Board (Rpm)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-3-4 Probe and MUX Board (RPM) The Probe and MUX (multiplexer) Board enables acquisition and processing of signals from and to probes connected to the front panel. The RPM Board has 192 receive/transmit channels, and performs the following functions: •...
  • Page 253: Back End Processor

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 5-4 Back End Processor 5-4-1 Introduction The Back End Processor (BEP) supports the operation of the Voluson i / Voluson e ultrasound unit and is the main controller for the system. The various components that comprise the Voluson i / Voluson e BEP are described in the following sections: •...
  • Page 254: Etx Sbc Central Processing Unit (Cpu)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-4-2 ETX SBC Central Processing Unit (CPU) The ETX SBC Central Processing Unit (CPU), which is mounted on top of the ETX Base Board, controls and processes the internal Back End Processor operations. In addition, the CPU supports the Front End via the PCI bus.
  • Page 255: Lcd Interface Block

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-4-3-2 LCD Interface Block The interface to the LCD is via the LCD JILI interface, designed to provide the following: A generic interface to LCD panels Backlight illumination control Backlight ON/OFF control Supports LVDS interface 5-4-3-3...
  • Page 256: Hard Disk

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-4-7 Hard Disk The Hard Disk is the main storage device of the Voluson i / Voluson e ultrasound unit. The Voluson i / Voluson e 80 GB Hard Disk, which is controlled by the CPU via signals from the IDE Bus, is divided into 4 partitions: C: System partition: Operating System (Windows XP or Windows XP embedded), including all Windows settings...
  • Page 257: Fan Control

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-4-9 Fan Control All fans have 3 wires with fan speed control input and actual speed output. Only four fans provide actual output; the fifth fan provides only On/Off information. All Fans are powered by +12V. All 5 fans operate simultaneously and are controlled via one output on the Fan Controller.
  • Page 258: Rtc Battery

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-4-11 RTC Battery The RTC battery provides power to the ETX ‘Real time clock‘. 5-4-12 Power Management Block The Power Management block provides the functionality described below. • Supports Three Power Modes: Full performance Standby mode Shutdown mode...
  • Page 259: Backend Connectors

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-4-13 BackEnd Connectors Figure 5-9 View of the Voluson i / Voluson e Rear Panel 5-4-13-1 External BackEnd Connectors Table 5-4 External BackEnd Connectors Connector Description View LAN 10/100 Base-TX Ethernet network connector rear connector panel SVGA Output (VCR option or CRT monitor option) rear connectors panel...
  • Page 260: Power Supply Unit

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 5-5 Power Supply Unit 5-5-1 Introduction The Power Supply Unit (GPP) provides the Voluson i / Voluson e system with both Low Voltage and High Voltage power, and also enables recharging of the battery. The GPP Unit controls following main functions: Input switching stage –...
  • Page 261: Power Management

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-5-3 Power Management 5-5-3-1 Power Operating Modes The Power Supply unit supports the following operation modes: Description Mode • Front End is OFF Shut Down • BEP is OFF. Only the PMC (Power Manager Controller) is working (powered by 3.3V_STB [less than 2ma]) •...
  • Page 262: Dc Source Selector

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-5-4 DC Source Selector This circuitry selects the source of the power. When the AC DC converter input into the system is higher than 18V, the system will be powered by the AC DC converter. Otherwise, the selector will look for an alternative power source (the battery), expecting to receive between 16 and 12V DC.
  • Page 263: Rechargeable Battery Pack Assy (Gpa)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-5-5 Rechargeable Battery Pack Assy (GPA) The lithium ion rechargeable battery provides power to the Voluson i / Voluson e system whenever an AC power source is not available. The system is supplied with a lithium ion rechargeable battery pack (GPA) installed in the battery bay, as standard.
  • Page 264 • If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal during use, recharging or storage, immediately remove it and stop using it. • If you have any questions about the battery, consult GE or your local representative. Storage of battery pack: Short term (less than one month): 0°...
  • Page 265: Battery Charger

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-5-6 Battery Charger 5-5-6-1 Introduction The system is provided with an internal and an optional external battery charger. The internal Battery Charger is capable of charging the battery pack with up to 0.5C (~3.3A) while keeping the overall dissipated power below a rated power.
  • Page 266: Recalibration

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-5-6-2 Recalibration The fuel gauge in the battery uses a highly accurate voltmeter, ammeter and time clock to measure actual charge in and out of the battery pack. In addition, algorithms are built into the smart-battery circuitry to compensate for the effects of discharge rate, discharge temperature, self-discharge and charging efficiency, etc.
  • Page 267: Keyboard And Control Panel

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 5-6 Keyboard and Control Panel Figure 5-12 Voluson i / Voluson e - Control Console Chapter 5 - Components and Functions (Theory) 5-35...
  • Page 268 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 5-7 Voluson i / Voluson e - key codes key code Description Voluson i / Voluson e - Functionality function depends on currently selected mode 1(1) Soft-menu button - function depends on currently selected mode 1(4) Soft-menu button ↑...
  • Page 269 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 5-7 Voluson i / Voluson e - key codes key code Description Voluson i / Voluson e - Functionality displays a cursor for menu and image operations Trackball pointer Dual-Screen format (vertical distribution) Dual Format (V) Quad-Screen format...
  • Page 270: Keyboard And Control Panel Layout

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-6-1 Keyboard and Control Panel Layout Figure 5-13 below illustrates the layout of the Voluson i / Voluson e keyboard and control panel. The buttons and controls are grouped together for ease of use; the different areas of the keyboard and control panel include the following: Soft-menu Buttons –...
  • Page 271: Keyboard And Control Panel Components

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-6-2 Keyboard and Control Panel Components 5-6-2-1 Power ON/OFF Button The Power On/Off button receives 3.3V from the BEP. Once shorted, this voltage is routed to the PMC which will activate or perform the appropriate task, as required. 5-6-2-2 Lid ON/OFF Switch The Lid On/Off switch is activated by a magnet located in the LCD frame (on the left side).
  • Page 272: Gain Rotary (Active Mode Button)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-6-2-5 Gain Rotary (Active Mode button) The Gain Rotary (located below the TGC Sliders - see Figure 5-13 on page 5-38) has a push button at its center. This is used to toggle between Active and Inactive mode. 5-6-2-6 Extended Keyboard The Extended keyboard (see...
  • Page 273: Display Module And Lcd Monitor

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 5-7 Display Module and LCD Monitor The Voluson i / Voluson e´s Display Module (GPD) consists of the 15” LCD Monitor in XGA (Extended Graphic Array) technology with built-in backlight lamp, flex connector board to the instrument, backlight inverter board, microphone (not available with SW 8.1.2 and higher), lid switch magnetic and the mechanical parts.
  • Page 274: Peripherals

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 5-8 Peripherals 5-8-1 Introduction The following peripherals are used with the Voluson i / Voluson e portable ultrasound system: • VCR (connected via VGA-to-video converter) • DVD recorder (connected via VGA-to-video converter) •...
  • Page 275: Dvd+(R)W Drive (Writer)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-8-1-3 DVD+(R)W Drive (Writer) The portable DVD+R/RW Drive (Writer) is used to backup images and reports. In addition, it is used as the main source of software upgrades and other service utility operations. The DVD+R/RW drive can read/write CDs and +DVDs.
  • Page 276: Air Flow Control

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 5-10 Air Flow Control 5-10-1 Air Flow Distribution The Voluson i / Voluson e cooling system includes five fans which are controlled by the PMC on the ETX baseboard. The fans suck air out of the system through the side vents, removing the warm air from around the internal components of the Voluson i / Voluson e - see Figure 5-15 below.
  • Page 277: Service Platform

    Local SERVICE • Access: This offers GE technicians the possibility to view the entire customer´s desktop Remote and operation system. Remote access to the Voluson i / Voluson e scanner requires permission and customer input to run diagnostics. 5-11-2-1 Local Access 1.) If not already in read mode, FREEZE the image.
  • Page 278: Remote Access

    Be sure the Voluson i / Voluson e is connected to the Internet and service platform is configured properly. If a GE Service technician requests remote access to your Voluson i / Voluson e scanner, activate the service platform as described below.
  • Page 279: Common Service Desktop (Csd)

    5-12-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. There are different possibilities to access the Common Service Desktop and its available features:...
  • Page 280: Error Logs

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-12-2 Error Logs When the Error Logs page is selected, different log viewing options are available. Log Viewer is displayed in a seperate window. Figure 5-20 Common Service Desktop - Error Logs 5-12-3 Diagnostics The Diagnostic page uses a web-controlled user interface to provide access to common service...
  • Page 281: Image Quality

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-12-4 Image Quality NOTE: This page is not populated in this version. Figure 5-22 Common Service Desktop - Image Quality 5-12-5 Calibration NOTE: This page is not populated in this version. Figure 5-23 Common Service Desktop - Calibration Chapter 5 - Components and Functions (Theory) 5-49...
  • Page 282: Configuration

    Figure 5-24 Common Service Desktop - Configuration Remote access is ONLY possible if the service platform is properly configured (either by the user NOTICE or a GE technican at site). Operation see: Section 7-5-3 "CSD: Configuration" on page 7-13. !! NOTICE:...
  • Page 283: Replacement

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-12-8 Replacement NOTE: This page is not populated in this version. Figure 5-26 Common Service Desktop - Replacement 5-12-9 NOTE: This page is not populated in this version. Figure 5-27 Common Service Desktop - PM Chapter 5 - Components and Functions (Theory) 5-51...
  • Page 284: Service Page

    The Service Page contains specific software/hardware test modules, system setup, update, etc. for Voluson systems only. 5-13-2 Service Login Extended Access (for GE field engineers only) see: NOTICE Section 5-11-2-1 "Local Access" on page 5-45. 1.) Press the UTILITIES key on the keyboard. The menu area changes to the Utilities menu.
  • Page 285: Auto Tester File

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Activate Service Auto Tester Monitor Test Printer Service Desktop Update Export System Data Delete all patients Battery Calibration Keyboard Layout TCP/IP Buffersize Figure 5-29 Service Tools window with Software 8.x.x 5-13-2-1 Auto Tester File Autotest is a log function of customer activities.
  • Page 286: Update

    Remote access to the Voluson i / Voluson e system requires permission and customer input to run diagnostics. If a GE Service technician requests remote access to your Voluson i / Voluson e scanner, activate the service platform as described in Section 5-11-2-2 on page 5-46.
  • Page 287: Delete All Patients

    Installation of printers or editing their settings is possible without entering the Windows Desktop. Operation see: 3-7-4 "Printer Installation manually" on page 3-87. Only accessories explicitly recognized and released by the system manufacturer GE WARNING WARNING may be used in connection with the system.
  • Page 288 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 2.) Select the Check and/or choose correct probe name from the pull down menu. DO NOT assign wrong coding (e.g., programming a RAB to a RIC probe). CAUTION An inaccurate probe programming will mostly likely damage the probe and/or system. check and/or choose probe form pull down menu check and/or enter...
  • Page 289: Boot Screen Functions

    Boot Screen Functions 5-14-1 Overview Following LINUX supported functions are available as soon as the “Boot Screen” appears: Voluson GE-Service Rollback Memtest Figure 5-32 Boot screen After 3 sec. without pressing any key, the system will boot-up in windows (= Voluson item).
  • Page 290: Rollback

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-14-1-3 Rollback This function offers the possibility to simply restore the previously used system configuration (rollback), which was stored on R:\ during “FMI from DVD”. 1.) Turn system OFF and then back ON. 2.) As soon as the “Boot Screen”...
  • Page 291: Memory Check In Linux

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 5-14-2 Memory Check in LINUX 1.) Turn system OFF and then back ON. 2.) As soon as the “Boot Screen” appears (see: Figure 5-32 on page 5-57), press the PG DN [↓...
  • Page 292 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 5-60 Section 5-14 - Boot Screen Functions...
  • Page 293: Overview

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 This chapter describes how to test and adjust the mechanical capabilities of a scanner that may be out of specification.
  • Page 294: Lcd Display (Monitor) Adjustment

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 6-3 LCD Display (Monitor) Adjustment 6-3-1 Color Calibration 1.) Press the UTILITIES key on the keyboard and then select the SYSTEM item. 2.) Select the page, enter the password and click the ACCEPT button.
  • Page 295: Lcd Brightness

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 6-3-2 LCD Brightness Adjusting the LCD brightness is one of the most important factors for proper image quality. If this is set incorrectly, the Gain, TGC, Dynamic Range and even Acoustic Output may have to be changed more often than necessary to compensate.
  • Page 296: Change Color Profil

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 6-3-3 Change Color profil 1.) Press the UTILITIES key on the alphanumeric keyboard. 2.) Select the SYSTEM item from the Utilities menu on the left side of the screen. 3.) Select the SERVICE page.
  • Page 297: Network Speed Settings

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 6-4 Network Speed Settings 1.) Press the UTILITIES key on the keyboard once to display the Utilities menu. 2.) Select SYSTEM and then open the page on the System Setup desktop screen. NETWORK 3.) Click the NETWORK ADAPTER CONFIGURATION button, read the message and confirm with YES.
  • Page 298: Modo Cart Height Adjustment

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 6-5 Modo Cart Height Adjustment Only adjust the height of the Modo Cart without the Voluson i / Voluson e and CAUTION any peripherals mounted. !! CAUTION: NOTE: A small 4mm hex wrench is needed to loosen the 2 screws on the rear of the Modo Cart.
  • Page 299: Dock Cart Height Adjustment

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 6-6 Dock Cart Height Adjustment Only adjust the height of the Dock Cart without the Voluson i / Voluson e and CAUTION any peripherals mounted. !! CAUTION: 1.) Pull the lever on the left side of the Dock Cart upwards to adjust the height of the Dock Cart. The top shelf will raise automatically.
  • Page 300: Dock Cart Light Height Adjustment

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 6-7 Dock Cart light Height Adjustment Only adjust the height of the Dock Cart light without the Voluson i / Voluson e and CAUTION any peripherals mounted. !! CAUTION: Be careful not to squeeze your fingers when adjusting the height of the Dock Cart light.
  • Page 301: Voluson Station Height Adjustment

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 6-8 Voluson Station Height Adjustment Only adjust the height of the Voluson Station without the Voluson i / Voluson e and CAUTION any peripherals mounted. !! CAUTION: Be careful not to squeeze your fingers when adjusting the height of the Voluson Station. CAUTION !! CAUTION: 1.) Pull the lever on the rear side of the Cart upwards to adjust the height of the Voluson Station.
  • Page 302: Modification Of Keyboard Layout

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 6-9 Modification of Keyboard Layout NOTE: Configuring the layout of the keyboard is only useful by changing the concerned keys also. see: Section 8-7 "Replacement of Key Caps (by special native language keys)" on page 8-25.
  • Page 303 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 6-9 Modification of Keyboard Layout (cont’d) 6.) Select the default input language from the drop down menu. default input language (drop down menu) corresponding keyboard layout [Add] button [Remove] button to remove unused language(s)
  • Page 304: Daylight Saving Time (Dst) - New Dates

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 6-10 Daylight Saving Time (DST) - New Dates The automatic Daylight Saving Time feature in Microsoft Windows products may no longer run on the Issue: correct date, causing the system’s time to be incorrect. Some countries have changed their start and end dates for Daylight Saving Time (DST).
  • Page 305 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
  • Page 306: Collect Vital System Information

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 7-2 Collect Vital System Information The following nformation is necessary in order to properly analyze data or images being reported as a malfunction or being returned to the manufacturer: Press the UTILITIES key, select the SYSTEM item and then enter the page.
  • Page 307: Needed Data - In Case Of System Crash Or Errors

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 7-2-1 Needed Data - in case of System crash or Errors In case of a system crash, or errors that are probably software bugs, provide as much detail information NOTICE as possible before contacting the OLC (On-Line Center).
  • Page 308: Shortcuts List

    Press the CTRL + H key simultaneous to display the shortcuts list and a description of what they do. NOTE: Figure 7-2 Shortcut list (Ctrl + H) For some shortcuts a service dongle and the “Standard GE” revolving password is required. After enabling “Extended Access”, press the LEFT TRACKBALL KEY once, followed by the shortcut.
  • Page 309 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Shortcut Description Ctrl + C copy text export system state (Full Backup and Dump files) to Alt + D D:\export Ctrl + H shortcuts list (see: Figure 7-2, above) H during start start UISApp in a mode ignoring all Ultrasound hardware Ctrl + Alt + I...
  • Page 310: Check Points Voltages

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 7-3 Check Points Voltages 7-3-1 How to check System Status Indicator LEDs There are two LEDs provided for Indications of Battery and external AC/DC source. left LED right LED AC Adapter Status Battery Status Figure 7-3 Status Indicator LEDs on Voluson i / Voluson e...
  • Page 311: Led Diagnose Patterns

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 7-3-2 LED diagnose patterns If an error condition will be recognized by the Power Management Controller SW, it will indicate the error by a special blinking code using the 2 power status leds (see Figure 7-3 on page 7-6).
  • Page 312: Screen Captures And Logs

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 7-4 Screen Captures and Logs There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center. This is accomplished by saving the image(s): A.) to SonoView and export them (as jpg, bmp or tiff) to DVD/CD+(R)W or external USB-drive.
  • Page 313: Export Log's And System Data

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 7-4-2 Export Log’s and System Data There are two possibilities to export system data (and log´s): 1.) by pressing the ALT + D key to save a snapshot and “Full Backup” of the System state; see: Section 7-4-2-1 on page 7-9 2.) via the EXPORT SYSTEM DATA button in the System Setup...
  • Page 314: Export Log's And System Data (Via Service Page)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 7-4-2-2 Export Log’s and System Data (via Service Page) 1.) Press the UTILITIES key on the keyboard, select the SYSTEM item from the menu area. 2.) Select the page on the screen. The “password window” appears automatically. SERVICE 3.) Enter the password and click the ACCEPT button to display the Service Tools window.
  • Page 315: Remote Access To Service Platform

    Be sure the Voluson i / Voluson e is connected to the Internet and service platform is configured properly in the Common Service Desktop - Configuration page. If a GE Service technician requests remote access to your Voluson i / Voluson e scanner, activate the service platform as described in Section 5-11-2-2 "Remote Access"...
  • Page 316 Figure 7-10 Common Service Desktop - Home a web-based user interface (UI) to provide access to common service components. The Service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface.
  • Page 317: Csd: Configuration

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 7-5-3 CSD: Configuration Figure 7-11 CSD - Configuration: Device Information / Advanced Configuration Figure 7-11, above shows the [Device Information] and [Advanced Configuration] tab in the Questra Agent Configuration menu. This pages normally do not have to be modified. 7-5-3-1 To configure Service Platform If a Proxy server is available, the information has to be properly entered, otherwise Questra...
  • Page 318: How To Use The Auto Tester Program

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 7-6 How to use the Auto Tester program NOTE: Startup the system. Before starting an exam Auto Tester must be turned on. Immediately after occurrence of the issue you need to stop the Auto Tester and save the results! In case of a system shut down, the Auto Tester stops automatically and the results will not be saved! 1.) Press the UTILITIES key and select the SYSTEM from the menu area on the left side of the screen.
  • Page 319 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 7-6 How to use the Auto Tester program (cont’d) 7.) Activate the “Auto Tester” program by clicking the “Record” icon on the displayed screen. “Record” icon Figure 7-15 Start Auto Tester 8.) Start scanning.
  • Page 320: Limitation Of The Auto Tester

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 7-6 How to use the Auto Tester program (cont’d) 9.) Maybe have it read to save the file, press the “Disc” icon, the following screen will appear: Figure 7-18 Auto Tester 10.)The file will be saved in D:\Export\AutotesterScripts\*.*.
  • Page 321: Minimum Configuration To Boot/Scan

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 7-7 Minimum Configuration to Boot/Scan 7-7-1 Booting without using the Ultrasound Hardware If the low-level ultrasound hardware is failing, the application can not be initialized correctly. Consequently, proper boot-up of the system is prevented. By steadily pressing the H key on the alphanumeric keyboard during startup, the application can be started, even if the ultrasound-hardware is broken.
  • Page 322 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 7-7-1 Booting without using the Ultrasound Hardware (cont’d) 7.) Save a “Full Backup” as described in Section 4-5-3 on page 4-32, or press the ALT + D key to gather detailed information and data from the system, see: Section 7-4-2-1 on page 7-9.
  • Page 323: Troubleshooting Trees And Instructions

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 7-8 Troubleshooting Trees and Instructions Table 7-6 below outlines the Voluson i / Voluson e troubleshooting trees and instructions shown in the sub-sections. Table 7-6 Troubleshooting Trees and Instructions Sub-section Description Page Number...
  • Page 324: Noise In 2D Mode

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 7-8-1 Noise in 2D Mode Figure 7-23 Noise in 2D Mode 7-20 Section 7-8 - Troubleshooting Trees and Instructions...
  • Page 325: Noise In Color Flow Mode

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 7-8-2 Noise in Color Flow Mode Figure 7-24 Noise in Color Flow Mode Chapter 7 - Diagnostics/Troubleshooting 7-21...
  • Page 326: Trackball - Impaired Sensitivity

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 7-8-3 Trackball - Impaired Sensitivity Figure 7-25 Trackball - Impaired Sensitivity 7-22 Section 7-8 - Troubleshooting Trees and Instructions...
  • Page 327: Printer Malfunction

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 7-8-4 Printer Malfunction Printer malfunction Is printer Check all signal and power supply cable properly connections between Printer and Scanner connected? Load paper by using the correct format. Is printer Check paper size (small, large etc.) loaded with paper? Check paper tray to fit with the used paper format...
  • Page 328 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 7-8-5 DVD/CD+R/RW Drive Malfunction Figure 7-27 DVD/CD+R/RW Drive Malfunction 7-24 Section 7-8 - Troubleshooting Trees and Instructions...
  • Page 329: Dvd/Cd+(R)W Drive Test

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 7-8-5-1 DVD/CD+(R)W Drive Test 1.) Insert an empty DVD/CD+(R)W into the Drive. 2.) Enter “Sonoview” by pressing the ARCHIVE key on the control panel; see Figure 7-28. 3.) Click the “Open” icon to display the list of exams. 4.) Select exam(s) and backup them to DVD/CD+(R)W.
  • Page 330: Network Troubleshooting

    IP address on the PC. If OK, the hardware connection inside the Scanner is OK. 7-8-6-2 GE remote service connection The customer gets visual information about the network status (shown in the status area on the right side of the screen). The network status is defined as follows: NO ICON - Cable disconnected or no network signal on a connected cable.
  • Page 331 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 7-9 Error Messages If the problem (error message) still exists after performing the described actions, call technical support. NOTICE Error Messages Actions AcousticUnitType not supported by BF_Interface: restart the system AcquMode - incorrect BBC_Wnd pointer restart the system Array bounds exceeded...
  • Page 332 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Error Messages Actions BF: can´t set C Rx Apod restart the system BF: can't set D RxApod restart the system binary file not found restart the system Binary path not set restart the system BM_Resample: overrun SampleLengthOnLineMem restart the system...
  • Page 333 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Error Messages Actions CRI: mismatch in steering angle header info restart the system CtrlMvGrpRepresentations::isInWriteMode not supported for restart the system mode nr xx CW Archive::not implemented restart the system CW quant TX Frequ out of range restart the system CW_BaseLinePos Out Of Range...
  • Page 334 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Error Messages Actions Electronic user manual not installed. Please install. install Electronic User Manual (EUM) and try it again End Bandwidth too big restart the system End ET too big restart the system End frequency too big restart the system...
  • Page 335 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Error Messages Actions File Type Unknown load volume files from other storage medium File Volume size not consistent load volume files from other storage medium FilterBuffer memory allocation failed restart the system FLT: Denormal operand restart the system...
  • Page 336 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Error Messages Actions incorrect VersionByte, xx of xx cycles OK contact technical support iPWCW_CloseProcessing in restart the system RT_PWorCWFilterBlock::execute() returned error code xx !! iPWCW_InitializeProcessing in restart the system RT_PWorCWFilterBlock::execute() returned error code xx !! iSetVideoSource(eVideoExtern) function fails restart the system...
  • Page 337 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Error Messages Actions Confirm the message by pressing the OK button and restart the system. Check functionality with a different 3D/ MotCtrl Unit: No probe reference position signal detected 4D probe.
  • Page 338 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Error Messages Actions Probe Scan Function Not Supplied restart the system PW Archive::not implemented restart the system PW_BaseLinePos Out Of Range press ok and save this user-setting once again PW_BurstCalcBlock: UserProgApplication out of range restart the system PW_CW_FFT_FactBlock: DSC_ScrollX_Zoom darf nicht...
  • Page 339 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Error Messages Actions RepresentationManager::vCreate: Nr. Dim 2 incorrect, see restart the system vSetReplayParams() call RepresentationManager::vResizeBufferLength called within restart the system Transaction!! RepresentationManager::vResizeBufferLength caused restart the system exception, replay buffer is not empty!! RepresentationManager::vSetGeoDescription called within restart the system Transaction!!
  • Page 340 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Error Messages Actions RT_MotionMBlock::execute caused exception:: Addr from DMA= xx, ReplayBuffAddr= xx, (Line+Header)Size= xx, restart the system blockLength= xx, value i= xx!! RT_PWorCW_Block::replay buffer size smaller than expected restart the system RT_PWorCW_Block::execute caused exception:: Addr from DMA= xx, ReplayBuffAddr= xx, (Line+Header)Size= xx, restart the system...
  • Page 341 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Error Messages Actions StoragePool: Not sufficient replay storage for M Mode available restart the system StoragePool: Not sufficient replay storage for MC or PW Mode restart the system available System detected severe error.
  • Page 342 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Error Messages Actions Unknown system exception restart the system unknown Xilinx-Version restart the system unrecordable disc or disc is closed try again with another disk Unsupported color mode restart the system Backup data are probably damaged.
  • Page 343 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter contains replacement procedures for different modules and their subsystems. The Manpower, time and Tools indicated in the Sub-sections include all requirements from NOTICE Preparations to Installation Procedures.
  • Page 344 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 8-1 Chapter 8 Contents Section Description Page Number Overview System Software - Installation/Upgrade Procedure 8-21 User Settings Only (Application Settings) Loading Procedure 8-21 Full Backup (Presets, Configurations & Appl. Settings) Loading Procedure 8-22 Software and Functional Checks after Installation/Upgrade Procedure 8-23...
  • Page 345: System Software - Installation/Upgrade Procedure

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-2 System Software - Installation/Upgrade Procedure 8-2-1 FMI from DVD 8-2-1-1 Introduction To update the Systems C:\ image via the FMI FROM DVD button in the System Setup page, NOTICE SERVICE...
  • Page 346 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Preparations (cont’d) 8-2-1-4 1.) Press the UTILITIES key on the keyboard once to display the Utilities menu. 2.) Select the SYSTEM item from the Utilities menu - which is displayed on the left side of the screen. 3.) Select the page on the System Setup desktop screen to see which Software/Hardware SYSTEM INFO...
  • Page 347 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Preparations (cont’d) 8-2-1-4 4.) Select the page on the System Setup desktop screen to see which Options are installed. OPTION D = Demo (Option is activated for demo and expires on the date shown in the “Valid”...
  • Page 348 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL If currently installed Application Software version is below 7.0.0: 1.) Press the USB button on the keyboard. 2.) In the displayed “Connect USB and Network Drives” window, select the MAP NETWORK DRIVE button to open the dialog.
  • Page 349: System Software - Installation Procedure (Fmi From Dvd)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-2-1-5 System Software - Installation Procedure (FMI from DVD) The system software installation procedure starts with saving and recording the settings present on the system (silent “Rollback”). Then the new software is written to the hard disk using the System DVD. Application Settings are automatically updated, to match with new Software version.
  • Page 350 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL System Software - Installation Procedure (FMI from DVD) (cont’d) 8-2-1-5 Figure 8-5 battery less charged or Power supply is disconnected In this case, connect the Power supply to charge the battery, afterwards try again. NOTE: To check capacity of the battery, move the cursor over the Battery Icon in the status area (see:...
  • Page 351 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL System Software - Installation Procedure (FMI from DVD) (cont’d) 8-2-1-5 NOTE: If the currently installed software has to be upgraded to a newer version, and the system DVD being used is SW 8.2.0 or higher, a new software specific “Permanent key is required.
  • Page 352 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL System Software - Installation Procedure (FMI from DVD) (cont’d) 8-2-1-5 15.)3 dots (one after the other) appear on the screen (see: Figure 8-11 below). 3 dots Figure 8-11 3 dots appear on the screen 16.)Booting auto..
  • Page 353 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL System Software - Installation Procedure (FMI from DVD) (cont’d) 8-2-1-5 17.)Updating will take some time..Figure 8-13 Please wait..18.)Please wait until all processes are finished (100 percent completed). 19.)The system is rebooting into windows (Boot screen - Voluson).
  • Page 354 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL System Software - Installation Procedure (FMI from DVD) (cont’d) 8-2-1-5 NOTE: If the BT warning dialog (Figure 8-15 below) appears, enter the Permanent key previously recorded 8-5) and then click SUBMIT. (refer to Figure 8-2: Option page (e.g., Voluson i) on page enter “Permanent key”...
  • Page 355: Fmi From External Device (Available At Software 8.X.x)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-2-2 FMI from external device (available at Software 8.x.x) 8-2-2-1 Introduction To update the Systems C:\ image via the FMI FROM EXTERNAL DEVICE button in the System Setup NOTICE page,see: Section 8-2-2-5 "System Software - Installation Procedure (FMI from external SERVICE...
  • Page 356 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Preparations (cont’d) 8-2-2-4 1.) Press the UTILITIES key on the keyboard once to display the Utilities menu. 2.) Select the SYSTEM item from the Utilities menu - which is displayed on the left side of the screen. 3.) Select the page on the System Setup desktop screen to see which Software/Hardware SYSTEM INFO...
  • Page 357 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Preparations (cont’d) 8-2-2-4 4.) Select the page on the System Setup desktop screen to see which Options are installed. OPTION D = Demo (Option is activated for demo and expires on the date shown in the “Valid”...
  • Page 358: System Software - Installation Procedure (Fmi From External Device)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-2-2-5 System Software - Installation Procedure (FMI from external device) The system software installation procedure starts with saving and recording the settings present on the system (silent “Rollback”). Then the new software is written to the hard disk using the Stick. Application Settings are automatically updated, to match with new Software version.
  • Page 359 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL System Software - Installation Procedure (FMI from external device) (cont’d) 8-2-2-5 Figure 8-19 battery less charged or Power supply is disconnected In this case, connect the Power supply to charge the battery, afterwards try again. NOTE: To check capacity of the battery, move the cursor over the Battery Icon in the status area (see:...
  • Page 360 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL System Software - Installation Procedure (FMI from external device) (cont’d) 8-2-2-5 14.)Check disk is performed automatically - restart. Figure 8-22 Check disk is performed automatically 15.)3 dots (one after the other) appear on the screen (see: Figure 8-23 below).
  • Page 361 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL System Software - Installation Procedure (FMI from external device) (cont’d) 8-2-2-5 16.)Booting auto..Figure 8-24 Boot screen - auto 17.)Updating will take some time..Figure 8-25 Please wait..18.)Please wait until all processes are finished (100 percent completed). Chapter 8 - Replacement Procedures 8-19...
  • Page 362 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL System Software - Installation Procedure (FMI from external device) (cont’d) 8-2-2-5 19.)The system is rebooting into windows (Boot screen - Voluson). Figure 8-26 automatic processes are running 20.)Please wait until all processes are finished. Finally the 2D screen is displayed on the monitor. NOTE: If the BT warning dialog (Figure 8-27...
  • Page 363: User Settings Only (Application Settings) Loading Procedure

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-3 User Settings Only (Application Settings) Loading Procedure 8-3-1 Introduction The User Settings contains: • User Programs • Auto Text • 3D/4D Programs 8-3-2 Loading Procedure see: Section 4-5-2 "Load User Settings Only (Application Settings)" on page 4-30 Section 8-4 Full Backup (Presets, Configurations &...
  • Page 364: Software And Functional Checks After Installation/Upgrade Procedure

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-5 Software and Functional Checks after Installation/Upgrade Procedure 1.) Press the UTILITIES key on the control panel once to display the Utilities menu. 2.) Select the SYSTEM item from the Utilities menu - which is displayed on the left side of the screen. 3.) Select to see which Software/Hardware version is installed in the unit.
  • Page 365: Replacement Or Activation Of Options

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-6 Replacement or Activation of Options • Following Options are available on Voluson i systems: Volume Ultrasound (3D Mode, Real Time 4D, VOCAL and TUI) VOCAL (seperat option on BT´09) DICOM®...
  • Page 366: Operation For Activating Options

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-6-1 Operation for activating Options 1.) Press the UTILITIES key on the keyboard to display the Utilities menu. 2.) Select the SYSTEM item from the menu area to invoke the setup desktop on the screen. 3.) Select the page where you can see which options are installed in the system.
  • Page 367: Replacement Of Key Caps (By Special Native Language Keys)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-7 Replacement of Key Caps (by special native language keys) Table 9-3 on page 9-5 shows the available Key Cap Kits. NOTICE Keys to be removed depend on the (special native) language kit. !! NOTICE: 8-7-1 Manpower...
  • Page 368: Replacement Of Tgc Slider Controls

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-8 Replacement of TGC Slider controls 8-8-1 Manpower One person, 5 minutes 8-8-2 Tools small-sized slotted screwdriver or tweezers 8-8-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 369: Replacement Of The Rotary Knob

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-9 Replacement of the Rotary Knob 8-9-1 Manpower One person, 5 minutes 8-9-2 Tools small-sized slotted screwdriver or tweezers 8-9-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 370: Replacement Of The Handle

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-10 Replacement of the Handle NOTE: The term "Handle" relates to the Voluson i / Voluson e portable carrying handle. 8-10-1 Manpower 10 minutes 8-10-2 Tools Use the appropriate flat and Phillips screw drivers as indicated in the Handle replacement procedure. 8-10-3 Preparations none...
  • Page 371: Replacement Of The Gpa - Battery

    !! WARNING: precautions listed in Voluson i / Voluson e Battery Safety on page 1-14. Use only batteries approved by GE as suitable for use with the Voluson i / Voluson e portable CAUTION ultrasound system. !! CAUTION: 1.) With the Voluson i / Voluson e lying face down (refer to...
  • Page 372: Replacement Of The Display Front Frame

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-12 Replacement of the Display Front Frame 8-12-1 Manpower One person, 15 minutes 8-12-2 Tools Phillips screwdriver 1 and 2 8-12-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 373: Replacement Of The Display Latches (Right/Left)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-13 Replacement of the Display Latches (right/left) 8-13-1 Manpower One person, 25 minutes 8-13-2 Tools Phillips screwdriver 1 and 2 8-13-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 374: Replacement Of The Bbdi Rigid Flex Assembly

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-14 Replacement of the BBDI Rigid Flex Assembly 8-14-1 Manpower One person, 40 minutes 8-14-2 Tools Phillips screwdriver 1 and 2 8-14-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 375 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-14-4 Rigid Flex Assy - Removal Procedure (cont’d) 2.) Using a small slotted screwdriver or your nails, release the locking “noses” at the Rigid Flex connector and then disconnect (pull out) the BBDI Rigid Flex. locking “noses”...
  • Page 376: Rigid Flex Assy - Installation Procedure

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-14-4 Rigid Flex Assy - Removal Procedure (cont’d) 5.) Disconnect the Microphone and the BBDI Inverter cable from the Flex (see: Figure 8-44 below). disconnect Micro cable disconnect BBDI Inverter cable disconnect LCD cable from Backlight Inverter board Figure 8-44 disconnect cables...
  • Page 377: Replacement Of The Lcd Backlight Inverter Board

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-15 Replacement of the LCD Backlight Inverter Board 8-15-1 Manpower One person, 25 minutes 8-15-2 Tools Phillips screwdriver 1 and 2 8-15-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 378: Replacement Of The Lcd Display

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-16 Replacement of the LCD Display 8-16-1 Manpower One person, 50 minutes 8-16-2 Tools Phillips screwdriver 1 and 2 8-16-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 379: Lcd Display - Installation Procedure

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-16-4 LCD Display - Removal Procedure (cont’d) 4.) Carefully lift the LCD Display upwards to gain access to the connector underneath. Remove the kapton tape from the LCD rear panel and disconnect and remove the BBDI Rigid Flex. remove kapton tape disconnect the BBDI Rigid Flex...
  • Page 380: Replacement Of The Display Rear Cover

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-17 Replacement of the Display Rear Cover 8-17-1 Manpower One person, 70 minutes 8-17-2 Tools Phillips screwdriver 1 and 2 8-17-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 381 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-17-4 Display Rear Cover - Removal Procedure (cont’d) 3.) Loosen and remove the 4 screws (2 on each of the left and right hinges); see: Figure 8-53 below. 2 screws on each of the left and right hinges Figure 8-53 loose and remove 4 screws at hinges 4.) Lift the Rear Cover away from the Bottom Assembly.
  • Page 382: Replacement Of The Modo Cart Components

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-18 Replacement of the Modo Cart Components For quick reference, the Modo Cart component replacement procedures are described in the following sub-sections: Table 8-2 Optional Peripherals and Accessories - Replacement Parts Sub-section Description Page Number...
  • Page 383: Replacement Of The Cable Clips

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-18-2 Replacement of the Cable Clips 8-18-2-1 Manpower One person, 5 minutes 8-18-2-2 Tools none 8-18-2-3 Preparations none 8-18-2-4 Cable Clips - Removal Procedure 1.) Pinch together the Cable Clips. 2.) Remove the Cable Clips.
  • Page 384: Replacement Of The Fuses

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-18-3 Replacement of the Fuses 8-18-3-1 Manpower One person, 10 minutes 8-18-3-2 Tools Phillips screwdriver 1 and 2 8-18-3-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72 8-18-3-4 Fuses - Removal Procedure 1.) Open the fuse protection on the rear of the isolation transformer.
  • Page 385: Replacement Of The Wheels

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-18-4 Replacement of the Wheels 8-18-4-1 Manpower Two persons, 20 minutes 8-18-4-2 Tools flat wrench 19 8-18-4-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72 2.) Put away all Peripherals Cables.
  • Page 386: Direction Kti106052, Revision 10 Service Manual

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-19 Replacement of the Dock Cart Components For quick reference, the Dock Cart component replacement procedures are described in the following sub-sections: Table 8-3 Optional Peripherals and Accessories - Replacement Parts Sub-section Description Page Number...
  • Page 387: Preparations

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-1 Replacement of the Probe Holder Kit 8-19-1-1 Manpower One person, 5 minutes 8-19-1-2 Tools none 8-19-1-3 Preparations none 8-19-1-4 Probe Holder Kit - Removal Procedure 1.) Pinch together the Probe Holders. 2.) Remove the Probe Holders.
  • Page 388: Preparation

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-2 Replacement of the Tray upper 8-19-2-1 Manpower One person, 10 minutes 8-19-2-2 Tools Phillips screwdriver 1 Allen wrench 5 8-19-2-3 Preparation 1.) Put away all things on the Tray (e.g. Probes) 8-19-2-4 Tray upper - Removal Procedure 1.) Loosen the 4 Allen screws that fasten the Tray upper and carefully remove it.
  • Page 389: Tray Lower - Removal Procedure

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-3 Replacement of the Tray lower 8-19-3-1 Manpower One person, 10 minutes 8-19-3-2 Tools Phillips screwdriver 1 Allen wrench 5 8-19-3-3 Preparation 1.) Put away all things on the Tray (e.g. Printer). 8-19-3-4 Tray lower - Removal Procedure 1.) Loosen the 4 Allen screws that fasten the Tray lower and carefully remove it.
  • Page 390: Replacement Of The Cover Bottom

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-4 Replacement of the Cover Bottom 8-19-4-1 Manpower One person, 10 minutes 8-19-4-2 Tools Phillips screwdriver 1 8-19-4-3 Preparations none 8-19-4-4 Cover Bottom - Removal Procedure 1.) For unlocking, turn the 2 quick release devices below the Cover Bottom 90°. quick release devices Figure 8-65 turn quick release devices 90°...
  • Page 391: Replacement Of The Back Column Cover

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-5 Replacement of the Back Column Cover 8-19-5-1 Manpower One person, 15 minutes 8-19-5-2 Tools Phillips screwdriver 1 8-19-5-3 Preparations 1.) Remove the Cover Bottom as described in Section 8-19-4 on page 8-48.
  • Page 392 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-6 Replacement of the Front Column Cover 8-19-6-1 Manpower One person, 20 minutes 8-19-6-2 Tools Phillips screwdriver 1 8-19-6-3 Preparation 1.) Remove the Tray upper as described in Section 8-19-2 on page 8-46.
  • Page 393: Replacement Of The Fuses

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-7 Replacement of the Fuses NOTE: Change Main Power Voltage and change fuses of Power Supply is the same procedure for the Power Supply of the Dock Cart and the Power Supply of the Voluson Station. 8-19-7-1 Manpower One person, 10 minutes...
  • Page 394: Replacement Of The Wheels

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-8 Replacement of the Wheels 8-19-8-1 Manpower Two persons, 20 minutes 8-19-8-2 Tools flat wrench 19 8-19-8-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 395 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-9 Replacement of the Mounting Platform 8-19-9-1 Manpower One person, 20 minutes 8-19-9-2 Tools Phillips screwdriver 1 Torx screwdriver TX8 8-19-9-3 Preparation 1.) Power OFF/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 396: Replacement Of The Tray Recorder

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-10 Replacement of the Tray Recorder 8-19-10-1 Manpower One person, 10 minutes 8-19-10-2 Tools Phillips screwdriver 1 8-19-10-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 397: Replacement Of The Probe Holder For Endocavity Probes

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-11 Replacement of the Probe Holder for Endocavity probes 8-19-11-1 Manpower One person, 5 minutes 8-19-11-2 Tools Phillips screwdriver 1 8-19-11-3 Preparations none 8-19-11-4 Probe Holder- Removal Procedure 1.) Screw off the 2 screws and carefully remove the Probe Holder. Figure 8-75 screw off the screws 8-19-11-5 Probe Holder- Installation Procedure...
  • Page 398: Replacement Of The Top Probe Holder

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-12 Replacement of the Top Probe Holder 8-19-12-1 Manpower One person, 2 minutes 8-19-12-2 Tools none 8-19-12-3 Preparations none 8-19-12-4 Top Probe Holder - Removal Procedure 1.) Carefully pull out the Top Probe Holder. Figure 8-76 pull out the Top Probe Holder 8-19-12-5 Top Probe Holder - Installation Procedure...
  • Page 399: Replacement Of The Probe Mux Box

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-19-13 Replacement of the Probe Mux Box 8-19-13-1 Manpower One person, 10 minutes 8-19-13-2 Tools Phillips screwdriver 1 Socket wrench 4 8-19-13-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 400: Probe Mux Box - Installation Procedure

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Probe Mux Box - Removal Procedure (cont’d) 8-19-13-4 3.) Screw off the 2 screws and the 1 Allen screw. unscrew 2 screws unscrew 1 Allen screw Figure 8-78 unscrew the 2 screws and the 1 Allen screw 4.) With both hands swing the Probe Box out away from the Dock Cart.
  • Page 401 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 8-20 Replacement of the Voluson Station Components For quick reference, the Voluson Station component replacement procedures are described in the following sub-sections: Table 8-4 Optional Peripherals and Accessories - Replacement Parts Sub-section Description Page Number...
  • Page 402: Replacement Of The Voluson Station Components

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-1 Replacement of the Cable Holder 8-20-1-1 Manpower One person, 5 minutes 8-20-1-2 Tools Phillips screwdriver 1 8-20-1-3 Preparations none 8-20-1-4 Cable Holder - Removal Procedure 1.) Screw off the 2 screws and carefully remove the Cable Holder. Figure 8-80 remove the Cable Holder 8-20-1-5 Cable Holder - Installation Procedure...
  • Page 403: Replacement Of The Probe Holder Insert

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-2 Replacement of the Probe Holder Insert 8-20-2-1 Manpower One person, 5 minutes 8-20-2-2 Tools none 8-20-2-3 Preparations none 8-20-2-4 Probe Holder Insert - Removal Procedure 1.) Pinch together the latches. Figure 8-81 pinch together the latches 2.) Remove the Probe Holder Insert.
  • Page 404: Direction Kti106052, Revision 10 Service Manual

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-3 Replacement of the Storage Tray Back 8-20-3-1 Manpower One person, 5 minutes 8-20-3-2 Tools Phillips screwdriver 1 8-20-3-3 Preparations none 8-20-3-4 Storage Tray Back - Removal Procedure 1.) Screw off the screw and carefully remove the Storage Tray Back. Figure 8-82 screw off the screw 8-20-3-5 Storage Tray Back - Installation Procedure...
  • Page 405: Replacement Of The Gel Condom Holder

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-4 Replacement of the Gel Condom Holder 8-20-4-1 Manpower One person, 5 minutes 8-20-4-2 Tools none 8-20-4-3 Preparations none 8-20-4-4 Gel Condom Holder - Removal Procedure 1.) Carefully pull out the Gel Condom Holder. Figure 8-83 pull out the Gel Condom Holder 8-20-4-5 Gel Condom Holder - Installation Procedure...
  • Page 406 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-5 Replacement of the Probe Holder for Endocavity Probes 8-20-5-1 Manpower One person, 5 minutes 8-20-5-2 Tools none 8-20-5-3 Preparations none 8-20-5-4 Probe Holder - Removal Procedure 1.) Carefully pull out the Probe Holder. Figure 8-84 pull out the Probe Holder 8-20-5-5 Probe Holder - Installation Procedure...
  • Page 407: Replacement Of The Additional Shelf Peripherals

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-6 Replacement of the Additional Shelf Peripherals 8-20-6-1 Manpower One person, 5 minutes 8-20-6-2 Tools Phillips screwdriver 1 8-20-6-3 Preparations 1.) Put away all things on the Shelf (e.g. Printer). 8-20-6-4 Additional Shelf Peripherals - Removal Procedure 1.) Screw off the 2 screws and carefully remove it.
  • Page 408: Replacement Of The Cover Bw Printer Shelf

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-7 Replacement of the Cover BW Printer Shelf 8-20-7-1 Manpower One person, 5 minutes 8-20-7-2 Tools Phillips screwdriver 1 8-20-7-3 Preparations none 8-20-7-4 Cover BW Printer Shelf - Removal Procedure 1.) Align tab on the front of the Printer Shelf Cover with the slot in front of the Printer Shelf;...
  • Page 409: Replacement Of The Console Bottom Shelf

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-8 Replacement of the Console Bottom Shelf 8-20-8-1 Manpower Oner person, 10 minutes 8-20-8-2 Tools Phillips screwdriver 1 8-20-8-3 Preparations 1.) Remove the Printer. 2.) Remove the Cover BW Printer Shelf as described in Section 8-20-7 on page 8-66.
  • Page 410 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-9 Replacement of the optional Probe Mux Box 8-20-9-1 Manpower One person, 5 minutes 8-20-9-2 Tools Phillips screwdriver 1 8-20-9-3 Preparation 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 411: Probe Mux Box - Installation Procedure

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Probe Mux Box - Removal Procedure (cont’d) 8-20-9-4 3.) With both hands swing the Probe Box out away from the Voluson Station. Figure 8-90 carefully remove the Probe Box 8-20-9-5 Probe Mux Box - Installation Procedure 4.) Fasten the Box on the Voluson Station with the screws as shown in...
  • Page 412: Replacement Of The Patient Monitor

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-10 Replacement of the Patient Monitor 8-20-10-1 Manpower One person, 5 minutes 8-20-10-2 Tools Phillips screwdriver 1 8-20-10-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 413: Replacement Of The Cover Bottom Voluson Station

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-11 Replacement of the Cover Bottom Voluson Station 8-20-11-1 Manpower One person, 10 minutes 8-20-11-2 Tools Phillips screwdriver 1 8-20-11-3 Preparations none 8-20-11-4 Cover Bottom - Removal Procedure 1.) Slide the Power Supply Bottom Cover backward and it will drop down. 2.) For unlocking, turn the 3 quick release devices below the Base Cover Assy 90°.
  • Page 414: Replacement Of The Top Shelf Additional Bottom Cover

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-12 Replacement of the Top Shelf Additional Bottom Cover 8-20-12-1 Manpower One person, 5 minutes 8-20-12-2 Tools Phillips screwdriver 1 8-20-12-3 Preparations none 8-20-12-4 Top Shelf Additional Bottom Cover - Removal Procedure 1.) Screw off the 4 screws and carefully remove it.
  • Page 415: Replacement Of The Top Shelf Bottom Rear Cover

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-13 Replacement of the Top Shelf Bottom Rear Cover 8-20-13-1 Manpower One person, 10 minutes 8-20-13-2 Tools Phillips screwdriver 1 8-20-13-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 416: Replacement Of The Top Column Back Cover

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-14 Replacement of the Top Column Back Cover 8-20-14-1 Manpower One person, 15 minutes 8-20-14-2 Tools Phillips screwdriver 1 8-20-14-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 417 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-15 Replacement of the Bottom Column Back Cover 8-20-15-1 Manpower Oner person, 15 minutes 8-20-15-2 Tools Phillips screwdriver 1 8-20-15-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 418: Replacement Of The Column Label Cover

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-16 Replacement of the Column Label Cover 8-20-16-1 Manpower One person, 15 minutes 8-20-16-2 Tools Phillips screwdriver 1 8-20-16-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 419 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-17 Replacement of the Top Column Front Cover 8-20-17-1 Manpower One person, 20 minutes 8-20-17-2 Tools Phillips screwdriver 1 8-20-17-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 420 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-17-5 Top Column Front Cover - Installation Procedure 1.) Place the new Top Column Front Cover and fasten it with the 4 screws. 2.) Mount the Bottom Column Back Cover as described in Section 8-20-15-5 on page 8-75.
  • Page 421 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-18 Replacement of the Bottom Column Front Cover 8-20-18-1 Manpower One person, 20 minutes 8-20-18-2 Tools Phillips screwdriver 1 8-20-18-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 422: Replacement Of The Top Shelf Bottom Front Cover

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-19 Replacement of the Top Shelf Bottom Front Cover 8-20-19-1 Manpower One person, 10 minutes 8-20-19-2 Tools Phillips screwdriver 1 8-20-19-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 423: Replacement Of The Front And Rear Wheels

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-20 Replacement of the Front and Rear Wheels 8-20-20-1 Manpower Two person, 20 minutes 8-20-20-2 Tools flat wrench 19 8-20-20-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 424: Replacement Of The Back Handle Complete

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-21 Replacement of the Back Handle complete 8-20-21-1 Manpower One person, 10 minutes 8-20-21-2 Tools Phillips screwdriver 1 8-20-21-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 425: Replacement Of The Power Supply Patient Monitor

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-22 Replacement of the Power Supply Patient Monitor 8-20-22-1 Manpower One person, 5 minutes 8-20-22-2 Tools none 8-20-22-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 426: Replacement Of The Image Resizer Box

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-23 Replacement of the Image Resizer Box 8-20-23-1 Manpower One person, 10 minutes 8-20-23-2 Tools Phillips screwdriver 1 8-20-23-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 427: Replacement Of The Lan Connector

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-24 Replacement of the LAN Connector 8-20-24-1 Manpower One person, 10 minutes 8-20-24-2 Tools Phillips screwdriver 1 8-20-24-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 428: Replacement Of The Vga Connector

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-25 Replacement of the VGA Connector 8-20-25-1 Manpower One person, 10 minutes 8-20-25-2 Tools Phillips screwdriver 1 8-20-25-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 429 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 8-20-26 Replacement of the DVD Drive 8-20-26-1 Manpower One person, 10 minutes 8-20-26-2 Tools Phillips screwdriver 1 8-20-26-3 Preparations 1.) Power Off/Shutdown the system as described in Section 3-6-3 on page 3-72.
  • Page 430: Direction Kti106052, Revision Service Manual

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 8-88 Section 8-20 - Replacement of the Voluson Station Components...
  • Page 431: Renewal Parts

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of replacement parts available for the Voluson i / Voluson e. NOTE: Furthermore refer to SN79040, which is the best source for new FRU parts that are not yet implemented in this Service Manual.
  • Page 432: List Of Abbreviations

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-2 List of Abbreviations Alternating Current Analog to Digital Converter Application Specific Integrated Circuit ASIC Assembly Assy Back End Processor Central Processing Unit Customer Replaceable Unit Common Service Desktop Control Ctrl Digital to Analog Converter...
  • Page 433 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-3 Parts List Groups Modo Cart Dock Cart Voluson Station Figure 9-1 Cart’s View Table 9-2 Mechanical and user accessible parts Item Part Group Name Table Number Description 100- Table 9-3 on page 9-5 Mechanical Hardware Parts (such as Fan and Speaker...
  • Page 434: Mechanical Hardware Parts And Covers

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-4 Mechanical Hardware Parts and Covers 111 112 113 114 115 116 117 118 119 122 123 124 125 126 127 128 129 Figure 9-2 mechanical Hardware Parts and Covers NOTE: The illustrations may not correspond to the actual product! Section 9-4 - Mechanical Hardware Parts and Covers...
  • Page 435 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-3 mechanical Hardware Parts and Covers Item Part Name Part Number Description KTZ154743 Bottom Cover (Bottom Housing) GPB1-1a - Bottom Basic Assy Bottom Cover without hinges KTZ302046 GPB10 Bottom Basic Assy (Bottom Housing - no PMCIA Slot) KTZ154757 Display Rear Cover (Top Housing) incl.
  • Page 436 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-3 mechanical Hardware Parts and Covers Item Part Name Part Number Description KTZ154797 kit (Rafi lasered, Whanam) incl. special native Key Cap Kit 2 - Spanish H48661CG language keys + tool for pulling out KTZ154798 kit (Rafi lasered, Whanam) incl.
  • Page 437: Keyboard Parts

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-5 Keyboard parts Figure 9-3 Keyboard parts NOTE: The illustrations may not correspond to the actual product! Chapter 9 - Renewal Parts...
  • Page 438 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-4 Keyboard Parts Item Part Name Part Number Description KTZ280191 Trackball GPU80 (incl. trackball ring) (Whanam) Trackball GPU80 (W) GPU80 Keyboard Module (Whanam) GPU80 Keyboard Module (W) KTZ301928 Voluson i / Voluson e Voluson i / Voluson e KTZ301972...
  • Page 439 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-6 Electronic Parts Figure 9-4 Electronic Parts NOTE: The illustrations may not correspond to the actual product! Chapter 9 - Renewal Parts...
  • Page 440 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-5 Electronic Parts Item Part Name Part Number Description AC Adapter (90 to 264V) for all countries; KTZ209076 AC/DC Adapter (for proper power cord see: Table 9-8) AC Adapter (90 to 264V) inverse polarity protection for AC/DC Adapter KTZ220798 all countries;...
  • Page 441 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-5 Electronic Parts Item Part Name Part Number Description Board RBT1.P1 LCD Backlight inverter Board RBT1.P1 KTZ301526 (only used for GPD5 and GPD6) LCD Backlight inverter KTZ301524 Board RBD2.P3 Flex board (use for GPD5 and GPD6) Board RBD2.P3 Flex board Board RBD3.P4 Flex Board KTZ303388...
  • Page 442: Electronic Parts

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-6 Electronic Parts (cont’d) 222 223 224 225 232 233 234 235 238 239 240 241 242 226 227 228 229 230 231 236 237 Figure 9-5 Electronic Parts NOTE: The illustrations may not correspond to the actual product! 9-12...
  • Page 443 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-6 Electronic Parts Item Part Name Part Number Description KTZ220334 ETX Base Board incl. Central Processing Unit (CPU) GPC1-1a - BackEnd (PC) Assy ETX Base Board incl. Central Processing Unit (CPU) KTZ220640 (can replace GPC1, SW 6.0.2 or higher is required and GPC2-2a - BackEnd (PC) Assy...
  • Page 444 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-6 Electronic Parts Item Part Name Part Number Description GPP6 DC/DC Power Supply KTZ302683 GPP6 DC/DC Power Supply (can replace GPP2, GPP3, GPP4 and GPP5) 9-14 Section 9-6 - Electronic Parts...
  • Page 445: Options And Upgrades

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-7 Options and Upgrades Table 9-7 Options and Upgrades Item Part Name Part Number Description bootable DVD for System HDD recovery Contents: SP with newest MS patches, Linux System/Boot DVD (6.0.2 ) KTZ196268 rescue partition, System C: Image, UISApp,...
  • Page 446 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-7 Options and Upgrades Item Part Name Part Number Description bootable DVD for System HDD recovery Contents: SP with newest MS patches, Linux System/Boot DVD (8.1.3) KTZ280151 rescue partition, System C: Image, UISApp, for Voluson i / Voluson e BT’09 Backup, EUM, Database Repair Tool, etc.
  • Page 447 BIOS Upgrade DVD for KTZ196370 Voluson i / Voluson e (ver. 890) Voluson i / Voluson e Voluson i BT06 toBT07 Upgrades H48661CY Upgrade kit contains: System DVD, BT’07 User Voluson i BT06 toBT07 Upgrades and Service Manual and Upgrade Manual...
  • Page 448 EVISION ERVICE ANUAL Table 9-7 Options and Upgrades Item Part Name Part Number Description Voluson i BT06/BT07/BT09 Voluson i BT06/BT07/BT09 H48681JS to 8.2.x upgrade L to 8.2.x upgrade L Voluson e BT06/BT07/BT09 Voluson e BT06/BT07/BT09 H48681JH to 8.2.x upgrade S to 8.2.x upgrade S...
  • Page 449 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL The options “RT 4D Biopsy” and “VCI” are applicable at Voluson i systems with BT´07 SW and higher. BT Version: BT-Version: The option “XTD-View” is only applicable at Voluson i / Voluson e with BT´07 software and higher. BT Version: BT-Version: The options “SonoVCAD heart”...
  • Page 450: Miscellaneous Cables

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-8 Miscellaneous Cables Table 9-8 Miscellaneous Cables Item Part Name Part Number Description Power Cord Europe 230V/240V KTZ220388 Power Cord - Europe 230V Power Cord USA Hospital Grade KTZ220389 Power Cord - USA Hosp.Grade Ca-Net Cable Brasilien 10A;2,5M...
  • Page 451 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-8 Miscellaneous Cables Item Part Name Part Number Description Power Cord China KTZ220391 Power Cord - China Power Cord Australia KTZ220392 Power Cord - Australia Power Cord India KTZ220387 Power Cord - India Power Cord United Kingdom (UK) 240V...
  • Page 452 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-8 Miscellaneous Cables Item Part Name Part Number Description Power Cord Argentina KTZ220478 Power Cord - Argentina Power Cord Israel KTZ220479 Power Cord - Israel Power Cord Switzerland KTZ220480 Power Cord - Switzerland Power Cord Denmark...
  • Page 453 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-8 Miscellaneous Cables Item Part Name Part Number Description Power Extender Cable for (optional) Peripherals, 1m (only used in combination with Modo Cart - H48651TD) KTZ220579 Power Extender Cable Power Extender Cable for (optional) Peripherals , 30cm (only used in combination with Modo Cart - H48651TD) KTZ220585...
  • Page 454 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-8 Miscellaneous Cables Item Part Name Part Number Description Cable Kit only used for Dock Cart (USB-Kabel (0,3m), Cabel Set DC Power Voluson i / Voluson eCart) KTZ280036 Cable Kit Network Cable only used for Dock Cart l KTZ280033...
  • Page 455 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-8 Miscellaneous Cables Item Part Name Part Number Description FLEX from RF-board (RBI) to Base Board KTZ220272 RBI to Base Board FLEX Base Board to Keyboard cable KTZ220331 Base Board to Keyboard cable SATA-Data cable 7pin,Power adapter cable 4pin, Converter-cable with USB-Power_Plug,...
  • Page 456: Optional Peripherals And Accessories

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-9 Optional Peripherals and Accessories Table 9-9 below outlines the optional peripherals and accessories described in the sub-sections. Table 9-9 Optional Peripherals and Accessories - Replacement Parts Sub-section Description Page Number 9-9-1...
  • Page 457: Printers

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-2 Printers Figure 9-7 Optional Peripherals and Accessories - Printers Chapter 9 - Renewal Parts 9-27...
  • Page 458 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-11 Optional Peripherals and Accessories - Printers Item Part Name Part Number Description KTZ220507 Digital B/W Video Printer Digital Black&White Video Printer (USB-Port) (H48651ML) (Sony UP-D897) KTZ302617 Mitsubishi B/W Printer P95D Mitsubishi B/W Printer P95D (H48681TH) Digital Color Printer...
  • Page 459: Drives & Additional Devices

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-3 Drives & additional Devices Figure 9-8 Optional Peripherals and Accessories - Drives & additional Devices Table 9-12 Optional Peripherals and Accessories - Drives & additional Devices Item Part Name Part Number Description portable DVD/CD Drive (Writer)
  • Page 460 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-3 Drives & additional Devices (cont’d) 530 531 532 533 Figure 9-9 Optional Peripherals and Accessories - Drives & additional Devices cont´d Table 9-13 Optional Peripherals and Accessories - Drives & additional Devices cont´d Item Part Name Part Number...
  • Page 461: Optional Equipment

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-4 Optional Equipment Figure 9-10 Optional Peripherals and Accessories - Optional Equipment NOTE: The illustrations may not correspond to the actual product! Chapter 9 - Renewal Parts 9-31...
  • Page 462 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-14 Optional Peripherals and Accessories - Optional Equipment Item Part Name Part Number Description KTZ196270 Scan/Freeze Footswitch USB only applicable at H48651T Voluson i / Voluson e systems with Footswitch BT Version software BT’07 and higher...
  • Page 463: Optional Modo Cart

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-5 Optional Modo Cart Figure 9-11 Modo Cart - Replacement parts Table 9-15 Modo Cart - Replacement parts Item Part Name Part Number Description KTZ196285 Modo Cart incl. isolation transformer, probe and gel Modo Cart (without peripherals) H48651TD holder, peripherals shelves, hand-rest and handles...
  • Page 464: Optional Dock Cart / Dock Cart Light

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-6 Optional Dock Cart / Dock Cart light Figure 9-12 Dock Cart / Dock Cart light - Replacement parts Table 9-16 Dock Cart / Dock Cart light - Replacement parts Item Part Name Part Number...
  • Page 465 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-6 Optional Dock Cart / Dock Cart light (cont’d) Figure 9-13 Dock Cart / Dock Cart light - Replacement parts (cont’d) Table 9-17 Dock Cart / Dock Cart light - Replacement parts (cont’d) Item Part Name Part Number...
  • Page 466 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-6 Optional Dock Cart / Dock Cart light (cont’d) Figure 9-14 Dock Cart / Dock Cart light - Replacement parts (cont’d) Table 9-18 Dock Cart / Dock Cart light - Replacement parts (cont’d) Item Part Name Part Number...
  • Page 467 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-6 Optional Dock Cart / Dock Cart light (cont’d) Figure 9-15 Dock Cart / Dock Cart light - Replacement parts (cont’d) Table 9-19 Dock Cart / Dock Cart light - Replacement parts (cont’d) Item Part Name Part Number...
  • Page 468: Optional Voluson Station

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-7 Optional Voluson Station Figure 9-16 Voluson Station - Replacement parts Table 9-20 Voluson Station - Replacement parts Item Part Name Part Number Description KTZ134656 Probe Cable Holder (1 pcs.) for Voluson E8 Cable Holder KTZ301238 Probe Holder Insert...
  • Page 469 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-7 Optional Voluson Station (cont’d) Figure 9-17 Voluson Station - Replacement parts (cont’d) Table 9-21 Voluson Station - Replacement parts (cont’d) Item Part Name Part Number Description KTZ280137 Column Cover complete Voluson Station Column Cover complete Voluson Station KTZ301228 Top Shelf Bottom Front Cover...
  • Page 470 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-7 Optional Voluson Station (cont’d) Figure 9-18 Voluson Station - Replacement parts (cont’d) Table 9-22 Voluson Station - Replacement parts (cont’d) Item Part Name Part Number Description KTZ301222 Power Supply Bottom Cover Power Supply Bottom Cover KTZ280145 Cover Bottom Voluson Station...
  • Page 471 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-7 Optional Voluson Station (cont’d) 619 620 617 618 Figure 9-19 Voluson Station - Replacement parts (cont’d) Table 9-23 Voluson Station - Replacement parts (cont’d) Item Part Name Part Number Description KTZ280134 Operating Panel Top incl.
  • Page 472 Probe Locker (4 pieces) KTZ280139 small parts kit Voluson Station screws for Probe Locker (4 pieces) Label GE Logo DM60mm (3 pieces) NOTE: The illustrations may not correspond to the actual product! 9-42 Section 9-9 - Optional Peripherals and Accessories...
  • Page 473 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-9-7 Optional Voluson Station (cont’d) Figure 9-21 Voluson Station - Replacement parts (cont’d) Table 9-25 Voluson Station - Replacement parts (cont’d) Item Part Name Part Number Description KTZ280149 Probe Mux Box Voluson Station Probe Mux Box Voluson Station H48681AL KTZ280142...
  • Page 474: System Manuals

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-10 System Manuals Table 9-26 System Manuals - Voluson i Part# / Kretz# Part# / Kretz# Part# / Kretz# Part# / Kretz# Part Name BT’06 BT’07 BT’09 BT’11 KTI106052 KTI106052 KTI106052...
  • Page 475 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-26 System Manuals - Voluson i Part# / Kretz# Part# / Kretz# Part# / Kretz# Part# / Kretz# Part Name BT’06 BT’07 BT’09 BT’11 H48661LA H48661BD H48671JD H48671JD Basic User Manual, Voluson i, Hungarian KTI106091 KTI106139...
  • Page 476 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-10 System Manuals (cont’d) Table 9-27 System Manuals - Voluson e Part# / Kretz# Part# / Kretz# Part# / Kretz# Part# / Kretz# Part Name BT’06 BT’07 BT’09 BT’11 KTI106052 KTI106052...
  • Page 477 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-27 System Manuals - Voluson e Part# / Kretz# Part# / Kretz# Part# / Kretz# Part# / Kretz# Part Name BT’06 BT’07 BT’09 BT’11 H48651YN H48681HF H48681HF Basic User Manual, Voluson e, Hungarian KTI106164 KTD100860 H48651YP...
  • Page 478 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-10 System Manuals (cont’d) Table 9-28 System Manuals - Voluson i/e CC200 Part# / Kretz# Part Name BT’14 KTI106052 Voluson i / Voluson e Service Manual System User Manuals H48691LU Basic User Manual, Voluson i/e, CC200, English H48691LW...
  • Page 479 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-28 System Manuals - Voluson i/e CC200 Part# / Kretz# Part Name BT’14 H48691MY Basic User Manual, Voluson i/e, CC200, Japanese H48691MZ Basic User Manual, Voluson i/e, CC200, Korean H48691NA Basic User Manual, Voluson i/e, CC200, Chinese H48691NB...
  • Page 480: Probes

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-11 Probes 9-11-1 2D-Probes Figure 9-22 2D Transducers NOTE: The illustrations may not correspond to the actual product! 9-50 Section 9-11 - Probes...
  • Page 481 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-29 2D Transducers Item Part Name Part Number Description 5131629 electronic broadband curved array transducer, frequency range of 2-5 4C-RS (H4000SR) MHz Applications: Abdominal, Obstetrics, Gynecology, Pediatric broadband curved array transducer, 4.0-9.0 MHz, 5434194 192 Elements,field of view: max.
  • Page 482: Real-Time 4D Volume Probes

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 9-11-2 Real-Time 4D Volume Probes Figure 9-23 Real-Time 4D Volume Probes Table 9-30 Real-Time 4D Volume Probes Item Part Name Part Number Description Real-time 4D broadband electronic curved-array transducer with a KTZ157007 frequency range of 2-5MHz.
  • Page 483: Biopsy Needle Guides

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 9-12 Biopsy Needle Guides Figure 9-24 Biopsy Needle Guides NOTE: The illustrations may not correspond to the actual product! Chapter 9 - Renewal Parts 9-53...
  • Page 484 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 9-31 Biopsy Needle Guides Item Part Name Part Number Description 12L-RS Multi- Non Sterile Multi Angle Bracket needle guide for probe 11L/12L-RS H40432LC angle bracket needle diameter: > 0.6 mm - < 2.1 mm **** Non Sterile 4C General Purpose Multi Angle Bracket E8385NA...
  • Page 485: Care & Maintenance

    10-1-1 Periodic Maintenance Inspections It has been determined by GE that your Voluson i / Voluson e system does not have any high wear components, therefore no Periodic Maintenance Inspections are mandatory. However, to maintain the safety and performance of the ultrasound system, a regular check by authorized personnel remains recommended.
  • Page 486: Why Do Maintenance

    It is the customer’s responsibility to ensure the Voluson i / Voluson e care & maintenance is performed as scheduled in order to retain its high level of safety, dependability and performance. Your GE Service Representative has an in-depth knowlegde of your Voluson i / Voluson e ultrasound scanning system and can best provide competent, efficient service.
  • Page 487 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-3-1 How often should care & maintenance tasks be performed? (cont’d) Table 10-2 Customer Care Schedule Item Service at Indicated Time Notes • AC Main Cable Inspect AC Main Cable Mobile Unit Check weekly Check if all cables are fixed well seated at also after corrective...
  • Page 488 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Table 10-2 Customer Care Schedule Item Service at Indicated Time Notes Check axial and lateral resolution (see Basic User Manual Technical specifications). Check Gain and TGC changes, vary the Also after corrective focus and check reaction on screen.
  • Page 489: Tools Required

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Special Tools, Supplies and Equipment used for Maintenance Table 10-3 Overview of Requirements for Care & Maintenance Tool Part Number Comments Digital Volt Meter (DVM) minimum 5% accuracy, 3.5 digit and 200 Ohm range required 46–194427P231 Kit includes anti–static mat, wrist strap and cables for 200 to 240...
  • Page 490: System Maintenance

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take ~15 minutes to perform. Refer to the system user documentation. Table 10-4 System Preliminary Checks Step Item Description Ask &...
  • Page 491: Peripheral/Option Checks

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-5-2-2 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to Table 3-13, “Approved Peripherals,” on page 3-103 for a list of approved peripherals.
  • Page 492: Physical Inspection

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-5-3 Physical Inspection Table 10-8 Physical Checks Step Item Description Labeling Verify that all system labeling is present and in readable condition. Scratches and Dents Inspect the system for dents, scratches or cracks. Verify ease of up/down movement and make sure that the LCD Display cover holds LCD Display Cover its set position when open and that there is no play on the mechanism.
  • Page 493: Cleaning

    The Basic User Manual and/or care card provides a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. Review the Basic User Manual of Voluson i / Voluson e for more details.
  • Page 494: Basic Probe Cleaning And/Or Disinfection

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-5-5-3 Basic Probe Cleaning and/or Disinfection Refer to the Basic User Manual of Voluson i / Voluson e for details on cleaning and desinfection. To help protect yourself from blood borne diseases, wear approved disposable gloves. WARNING WARNING These are made of nitrile derived from vegetable starch to prevent allergic latex...
  • Page 495: Disinfection By Means Of The Ric-Holder

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-5-5-4 Disinfection by means of the RIC-Holder Especially for Real-time 4D endocavity probes (RIC), it is necessary to take extreme care when transporting the system with the probe attached, or during the disinfection process. Inadequate handling may lead to dead elements, shocked head mechanics, etc.
  • Page 496: Electrical Safety Tests

    Compare all safety-test results with safety-test results of previously performed safety tests CAUTION (e.g., last year etc.). In case of unexplainable abrupt changes of safety-test results consult experienced authorized service personnel or GE for further analysis. Equipment users must ensure that safety inspections are performed whenever damage WARNING...
  • Page 497: Leakage Current Limits

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-7-2 Leakage Current Limits The following acceptance limits and test conditions are summarized from IEC60601 including national deviations and IEC 62353 and in some cases are lower than that specified by the standards. In accordance with these standards, fault conditions like Reverse Polarity of the supply mains and Open Neutral are no longer required for field evaluation of leakage current.
  • Page 498: Outlet Test - Wiring Arrangement - Usa & Canada

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-7-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated. Any problems found should be reported to the hospital immediately and the receptacle should not be used.
  • Page 499: Grounding Continuity (Systems With Modo Cart Only)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-7-4 Grounding Continuity (Systems with Modo Cart only) Electric Shock Hazard! CAUTION The patient or operator MUST NOT come into contact with equipment during this test. !! CAUTION: NOTE: This test is not valid for stand-alone Voluson i / Voluson e systems and is only applicable when the Voluson i / Voluson e portable ultrasound system is mounted on the Modo Cart.
  • Page 500: Grounding Continuity (Systems With Dock Cart Only)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-7-5 Grounding Continuity (Systems with Dock Cart only) Electric Shock Hazard! CAUTION The patient or operator MUST NOT come into contact with equipment during this test. !! CAUTION: NOTE: This test is not valid for stand-alone Voluson i / Voluson e systems and is only applicable when the Voluson i / Voluson e portable ultrasound system is mounted on the Dock Cart.
  • Page 501: Grounding Continuity (Systems With Voluson Station Only)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-7-6 Grounding Continuity (Systems with Voluson Station only) Electric Shock Hazard! CAUTION The patient or operator MUST NOT come into contact with equipment during this test. !! CAUTION: NOTE: This test is not valid for stand-alone Voluson i / Voluson e systems and is only applicable when the Voluson i / Voluson e portable ultrasound system is mounted on the Voluson Station.
  • Page 502: Definition

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-7-7 Chassis Leakage Current Test Electric Shock Hazard. DANGER When the meter's ground switch is OPEN, DO NOT touch the unit! !! DANGER: Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit CAUTION is powered ON.
  • Page 503: Data Sheet For Enclosure/Chassis Leakage Current

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-7-7-3 Data Sheet for Enclosure/Chassis Leakage Current Table 10-14 shows a typical format for recording the enclosure/chassis leakage current. Measurements should be recorded from multiple locations for each set of test conditions. The actual location of the test probe may vary by system.
  • Page 504: Isolated Patient Lead (Source) Leakage-Lead To Ground

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-7-8 Isolated Patient Lead (Source) Leakage–Lead to Ground Equipment damage possibility. Never switch the Polarity when the unit is powered ON. CAUTION Be sure to turn the unit power OFF before switching the polarity using the POLARITY switch. Otherwise, the unit may be damaged.
  • Page 505: Isolated Patient Lead (Source) Leakage-Lead To Lead

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-7-9 Isolated Patient Lead (Source) Leakage–Lead to Lead When using the Dale 600, switch the meter's function selector to the LEAD-LEAD position. Select and test each of the ECG lead positions (except ALL) on the LEAD selector, testing each to the power and ground condition combinations found in: •...
  • Page 506: Isolated Patient Lead (Sink) Leakage-Isolation Test

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-7-10 Isolated Patient Lead (Sink) Leakage-Isolation Test Line voltage is applied to the ECG leads during this test. To avoid possible electric shock CAUTION hazard, the system being tested must not be touched by patients, users or anyone while the ISO TEST switch is depressed.
  • Page 507: Probe Leakage Current Test

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL 10-7-11 Probe Leakage Current Test Do not use the probe if the insulating material has been punctured or otherwise DANGER compromised. Integrity of the insulation material and patient safety can be verified by safety testing according to IEC60601-1.
  • Page 508: When There's Too Much Leakage Current

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 10-8 When There's Too Much Leakage Current... 10-8-1 Chassis Fails Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
  • Page 509: Ultrasound Equipment Quality Check (Eqc)

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 10-9 Ultrasound Equipment Quality Check (EQC) Please download and use the latest version of this forms. They can be retrieved from MyWorkshop. • EQC - Refer to DOC0929340 in MyWorkshop •...
  • Page 510: Electrical Safety Test Form

    i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Section 10-10 Electrical Safety Test Form Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground & wiring config.) System Ground Continuity Chassis Source Leakage Current - Potential equilibrium connector Chassis Source Leakage Current - Monitor housing...
  • Page 511 Replacement Procedure, 8-55 Top Level Architecture, 5-8 Dock Cart - Probe Holder Kit Boot Screen Functions Replacement Procedure, 8-45 GE-Service, 5-57 Dock Cart - Probe Mux Box Memtest, 5-58 Replacement Procedure, 8-57 Rollback, 5-58 Dock Cart - Top Probe Holder...
  • Page 512 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Disconnection, 3-44 Measure Setup Software Configuration Checks, 4-38 Media Formatting, 4-27 FMI from DVD Models Covered, 1-2 Installation Procedure, 8-7 Modo Cart - Cable Clips Theory, 5-54 Replacement Procedure, 8-41 Formatting Modo Cart - Fuses Media, 4-27...
  • Page 513 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Bearing Handle, 8-28 How to change Language, 3-99 Display Front Frame, 8-30 How to enter Date and Time, 3-99 Display Latches (right/left), 8-31 How to enter Hospital Name, 3-99 Display Rear Cover, 8-38 System Maintenance, 10-6 Dock Cart - Back Column Cover, 8-49...
  • Page 514 i / V OLUSON OLUSON KTI106052, R IRECTION EVISION ERVICE ANUAL Index...
  • Page 516 2006 - 2013, General Electric Company. © GE Medical Systems, a General Electric Company, going to market as GE. GE Healthcare Austria GmbH & Co OG Tiefenbach 15 A-4871 Zipf Austria www.gehealthcare.com...