GE LOGIQ E9 Service Manual page 114

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5535208-100, R
IRECTION
Table 3-12
5.
6.
3 - 50
. 2
EV
Dataflow
Navigate to Utility -> Connectivity -> Button, select a print button and add "Copy to Dataflow"
to its Printflow View. DO NOT add the DICOM Image Storage Service to the Printflow as you
would run the risk of sending each image to the service twice. Set the other parameters of the
print button as you choose, using "Print Button Settings" in section
Settings" on page 3-52
as a guide; but keep in mind that the settings of the service will override
the settings of the button for the images sent to the service.
Patient Page - Dataflow Drop-down Menu
Back on the Patient page, select your dataflow from the drop-down menu. From that point on,
pressing your configured print button sends your images to the service(s) added to your
dataflow.
NOTE: Whenever you change the settings of a dataflow, before you use it, select a different
dataflow or "No Archive" on the Patient screen, and then reselect the changed dataflow. This
forces the to read the modified resource file and implement your changes. If you change the
settings of any service to which the dataflow directs traffic, you do not need to reselect it.
Section 3-9 - Connectivity Setup and Tips
->
->
Utility
Connectivity
LOGIQ E9 S
Button
Table 3-14 "Print Button
M
ERVICE
ANUAL

Advertisement

Table of Contents
loading

Table of Contents