GE
D
5535208-100, R
IRECTION
Table 8-60
4.
Lay the Monitor flat.
Remove the two lower Phillips screws (1).
Loosen the upper Phillips screws (2). The
Monitor will be supported by the two upper
screws.
Slide the Monitor off of the Arm Assembly.
Remove the upper Phillips screws to
reinstall into the replacement LCD Monitor.
8 - 98
. 2
EV
LCD Monitor V2 removal - R5.x and later
Steps
Section 8-6 - Replacing Top Console Parts
LOGIQ E9 S
Corresponding Graphic
Upper Monitor mounting screw and hole
M
ERVICE
ANUAL