Diagnostics Window Overview (cont'd)
The Service Diagnostics windows have some advanced features.
NOTE:
The Loop Count field must have a value of 2 or greater to activate the Advanced features.
8.) This button toggles between Standard and Advanced mode. The window below shows the
advanced features.
9.) Select the choice for Stop On Failure for the Looping and List sections:
-
-
10.)Select the choice for Stop On Abort for the Looping and List sections:
-
-
-
. 2
EV
Figure 7-24 Diagnostic Window Overview
Yes = the test will abort at the first failure
No = the test will complete even if there are failures
List = the list will abort when you click the Abort button
Loop = the looping will abort when you click the Abort button
Diag = the diagnostic test will abort when you click the Abort button
Chapter 7 Diagnostics/Troubleshooting